Step Number

Description

Part Number

1

1 mL Cryobank vials (96 tubes per rack)

374088

2

8 channel handheld screw cap capper/decapper with stand for

Thermo Scientific Nunc-style tubes

4105NUN

3

Thermo Scientific

Matrix

96 deep-well automation reservoir

1064-15-6

4

Filter plate

Chromsystems

2–8

Thermo Scientific Versette system

- Base unit

- 96 and 384 channel pipetting module

- 6 position stage

- 96 channel air disp. pipetting head, volume 5–300 µL

650-01-BS

650-02-NTC

650-03-SPS

650-06-96300

4–7

Tips – 300 µL extended-length D.A.R.T.’S

5536

8

Titer plate shaker

4625-1CEQ

9

Thermo Scientific

Heraeus

Labofuge

400 with microplate rotor package

75008371

10

Nunc pierceable 96 well cap mats

Nunc polypropylene microplate - 96 well (1.0 mL)

276002

260252

Table 1. Parts and consumables required for specific steps in the liquid handling procedure

Figure 2. Liquid handling procedure

1. The LIMS produces worklist, and

Cryobank vials containing samples

are organized in a 96 well rack (Figure 3).

2. Vials are decapped (8 channel

decapper) and placed on

Versette system (Figures 4 and 5).

3. Reagent reserviors are filled (IS and

precipitation reagent).

4. The empty filter plate and filled reagent

reservoirs are placed on the Versette

system. D.A.R.T.’S tips are replaced.

5. 100 µL sample is transferred to filter

plate. D.A.R.T.’S tips are replaced.

6. 200 µL IS is transferred to filter plate.

D.A.R.T.’S tips are replaced.

7. 25 µL IS precipitation reagent is

transferred to filter plate.

D.A.R.T.’S tips are discarded.

8. The filter plate is removed, covered and

mixed (plate shaker, 600 rpm, 10 min).

9. The filter plate is placed on top of

the microtitre collection plate and

centrifuged (2,000

g

, 4 min).

10. The collection plate is sealed and

placed on the autosampler for analysis

with the TurboFlowmethod and

LC-MS/MS (Figure 6).

4 minutes

Preparation of two

filter plates prior to

simultaneous mixing

and centrifugation

increases throughput.

Figure 3. Cryobank vials

Figure 5. Versette liquid handling system and

one rack of D.A.R.T.’S tips

Figure 4. 8 Channel

handheld screw cap

capper/decapper

Figure 6. Transcend system and TSQ Vantage mass spectrometer