AN63738_E 10/13S

Africa-Other

+27 11 570 1840

Australia

+61 3 9757 4300

Austria

+43 810 282 206

Belgium

+32 53 73 42 41

Canada

+1 800 530 8447

China

+86 10 8419 3588

Denmark

+45 70 23 62 60

Europe-Other

+43 1 333 50 34 0

Finland

+358 9 3291 0200

France

+33 1 60 92 48 00

Germany

+49 6103 408 1014

India

+91 22 6742 9494

Italy

+39 02 950 591

Japan

+81 45 453 9100

Latin America

+1 561 688 8700

Middle East

+43 1 333 50 34 0

Netherlands

+31 76 579 55 55

New Zealand

+64 9 980 6700

Norway

+46 8 556 468 00

Russia/CIS

+43 1 333 50 34 0

South Africa

+27 11 570 1840

Spain

+34 914 845 965

Sweden

+46 8 556 468 00

Switzerland

+41 61 716 77 00

UK

+44 1442 233555

USA

+1 800 532 4752

www.thermoscientific.com

©2013 Thermo Fisher Scientific Inc. All rights reserved. ISO is a trademark of the International Standards Organization.

All other trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is presented as an

example of the capabilities of Thermo Fisher Scientific products. It is not intended to encourage use of these products in any

manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change.

Not all products are available in all countries. Please consult your local sales representative for details.

Thermo Fisher Scientific,

San Jose, CA USA is

ISO 9001:2008 Certified.

0

50

100

150

200

250

300

350

400

0

5000 10000 15000 20000 25000 30000 35000 40000 45000 50000

Y = 8.337e-3X + 3.048e-1; R^2: 0.9946; Origin: Ignore; W: 1/X; Area

0

10

20

30

40

50

60

70

80

0

10000 20000 30000 40000 50000 60000 70000 80000 90000 100000

Y = 7.43e-4X + 1.18e-2; R^2: 0.9973; Origin: Ignore; W: 1/X; Area

EtG

Calibrator

%Recovery

25 ng/mL

-3.45

50 ng/mL

9.81

100 ng/mL

-3.28

250 ng/mL

3.96

500 ng/mL

4.39

1000 ng/mL

-12.4

5000 ng/mL

-10.1

10,000 ng/mL

12.4

50,000 ng/mL

-1.28

EtS

Calibrator

%Recovery

100 ng/mL

0.93

250 ng/mL

8.64

500 ng/mL

-11.8

1000 ng/mL

0.14

5000 ng/mL

2.41

10,000 ng/mL

9.65

50,000 ng/mL

-7.26

100,000 ng/mL

2.58

Application Note 589

Figure 2. Representative chromatograms showing quantifier and confirming ions for

EtG and EtS at 100 and 25 ng/mL, respectively

Figure 1. Representative calibration curves for EtG and EtS in urine

1.6

1.8

2.0

2.2

2.4

Time (min)

0

20

40

60

80

100

RelativeAbundance

0

20

40

60

80

100

RelativeAbundance

RT: 2.10

AA: 1742

SN:

329

RT: 2.10

AA: 889

SN:

131

0

20

40

60

80

100

RelativeAbundance

0

20

40

60

80

100

RelativeAbundance

RT: 2.75

AA: 8229

SN:

169

RT: 2.74

AA: 2444

SN:

249

2.5

2.6

2.7

2.8

2.9

3.0

3.1

3.2

Time (min)

EtG quanƟfier ion at 100 ng/mL

EtS quanƟfier ion at 25 ng/mL

EtG confirming ion at 100 ng/mL

EtS confirming ion at 25 ng/mL

Conclusion

• This method provides limits of detection in urine of

100 ng/mL for EtG and 25 ng/mL for EtS.

• SPE extraction helps remove interferences in urine.

• Addition of an ion-pairing reagent helps

chromatographically separate the compounds from

interferences on the solvent front, thereby improving

limits of detection.

• This method is suitable for forensic toxicology.

References

1.

LC/MS Method For Extracting Ethyl Glucuronides

From Urine Using: 200 mg Clean Screen

®

Extraction

Column;

Part #: CSETG203; UTC, Inc. Bristol, PA;

2008

2. Thermo Scientific Application Note 488b,

Quantitation of Urinary Ethyl Glucuronide and Ethyl

Sulfate Using Ultrahigh Resolution LC-MS, http://

planetorbitrap.com/data/uploads/ZFS1327868383151_

AN488b.pdf

For Forensic Toxicology Use Only.