The intraday and interday analytical accuracies and

variability are presented in Tables 4 and 5, respectively.

For each analyte, intraday variability and accuracy were

determined by analyzing five samples at each QC level.

For interday variability and accuracy, the process

performed for intraday data was repeated on three

different days. For the intraday study, the accuracy

obtained was between 94% and 112% with a %RSD

less than 12%. In the case of interday assays, the accuracy

was between 93% and 111% with a %RSD less than

13%. The obtained results show a low variability for the

two QC levels as well as good accuracy.

Figure 2. Chromatograms of the SRM transitions of some commonly found benzodiazepines at the LOQ