Application Note 588

Conclusion

Using the Prelude SPLC system, high-throughput, cost-

efficient solutions were developed for forensic analysis of

opiates and benzodiazepines in urine. The methods met

industry requirements for precision, accuracy, and

robustness. Implementation of the method on a Prelude

SPLC simplified the work flow and resulted in a 40–60%

reduction of solvent usage due to the ability of the system to

utilize high efficiency, small diameter columns. The mobile

phase volumes in developed methods were approximately

3.8 mL per sample, which reduced cost of reagents and

waste disposal. Multiplexing into a single mass spectrometer

increased MS utilization and reduced overall system

hardware costs relative to two independent LC-MS systems.

The Prelude SPLC system makes multiplexing of two

different methods, with or without on-line sample prep,

possible and enabled a throughput of 480 samples in

24 hours. The implementation of methods was facilitated by

the many ease-of-use features incorporated into the system.

Acknowledgement

Thanks to Kent Johnson, Director of Toxicology at

Pacific Hospital of Long Beach, CA for providing donor

urine samples and data collected with their validated

LC-MS method.

For Forensic Toxicology use only.

www.thermoscientific.com

©2013 Thermo Fisher Scientific Inc. All rights reserved. ISO is a registered trademark of the International Organization for Standardization

(Organisation Internationale De Normalization). All other trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This

information is presented as an example of the capabilities of Thermo Fisher Scientific products. It is not intended to encourage use of these

products in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change.

Not all products are available in all countries. Please consult your local sales representative for details.

Table 6. Results of matrix effect experiment showing percent recovery of benzodiazepines in spiked urine

Urine

Lot#

% Recovery

Desalkyflura-

zepam

Diazepam

Lorazepam

Midazolam

Nordiazepam

Oxazepam

Temezepam

1

114

118

110

117

110

105

111

2

109

111

104

116

112

99.8

105

3

108

112

103

117

113

103

102

4

107

114

108

114

118

105

106

5

105

115

108

117

112

99.0

106

6

114

113

109

115

111

104

105

7

108

113

96.2

117

112

98.4

103

8

107

107

101

112

111

95.7

103

Urine

Lot#

% Recovery

2-Hydroxy-

ethylflurazepam

7-Aminoclona-

zepam

7-Aminoflunitra-

zepam

7-Aminonitra-

zepam

a

-Hydroxy-

alprazolam

a

-Hydroxy-

triazolam

Alprazolam

1

112

104

103

99.2

111

114

116

2

102

99.8

103

104

112

116

113

3

106

103

101

102

113

116

108

4

112

104

106

100

108

111

114

5

100

102

102

95.8

110

111

108

6

118

105

109

104

113

118

111

7

106

101

99.7

104

111

124

110

8

107

97.8

98.5

101

112

93.0

107

AN63754_E 10/13S

Africa-Other

+27 11 570 1840

Australia

+61 3 9757 4300

Austria

+43 810 282 206

Belgium

+32 53 73 42 41

Canada

+1 800 530 8447

China

+86 10 8419 3588

Denmark

+45 70 23 62 60

Europe-Other

+43 1 333 50 34 0

Finland

+358 9 3291 0200

France

+33 1 60 92 48 00

Germany

+49 6103 408 1014

India

+91 22 6742 9494

Italy

+39 02 950 591

Japan

+81 45 453 9100

Latin America

+1 561 688 8700

Middle East

+43 1 333 50 34 0

Netherlands

+31 76 579 55 55

New Zealand

+64 9 980 6700

Norway

+46 8 556 468 00

Russia/CIS

+43 1 333 50 34 0

South Africa

+27 11 570 1840

Spain

+34 914 845 965

Sweden

+46 8 556 468 00

Switzerland

+41 61 716 77 00

UK

+44 1442 233555

USA

+1 800 532 4752

Thermo Fisher Scientific,

San Jose, CA USA is

ISO 9001:2008 Certified.