          Shown below is a decrease in retention for cytosine on the four stationary phases as the column temperature
        
        
          is increased. Linear van’t Hoff plots resulted from this study. Negative retention enthalpy values were obtained,
        
        
          indicating an exothermic process of transferring cytosine from the mobile phase to the stationery phase.
        
        
          0.0036
        
        
          ln k
        
        
          1/T (K)
        
        
          1.00
        
        
          10.00
        
        
          0.0030
        
        
          0.0032
        
        
          0.0034
        
        
          0.0028
        
        
          Syncronis HILIC
        
        
          Hypersil GOLD HILIC
        
        
          Accucore HILIC
        
        
          Hypersil GOLD Silica
        
        
          Independently of its effect on the retention of test compounds, column temperature often positively affects the
        
        
          chromatographic efficiency due to lower viscosity and higher diffusivity. Our study demonstrated lower back
        
        
          pressures could be achieved when the temperature was raised, with average pressure dropping from 113
        
        
          to 77 bar as the temperature was taken from 20 to 70 °C.
        
        
          0.0036
        
        
          ln k
        
        
          1/T (K)
        
        
          0.00
        
        
          0.0030
        
        
          0.0032
        
        
          0.0034
        
        
          0.0028
        
        
          0.01
        
        
          0.10
        
        
          1.00
        
        
          10.00
        
        
          Syncronis HILIC
        
        
          Hypersil GOLD HILIC
        
        
          Accucore HILIC
        
        
          Hypersil GOLD Silica
        
        
          Van't Hoff plot for salicylic acid
        
        
          Van't Hoff plot for cytosine
        
        
          Mobile Phase:
        
        
          90/10 acetonitrile/100 mM ammonium
        
        
          acetate
        
        
          Flow Rate:
        
        
          1.0 mL/min
        
        
          Inj.Volume:
        
        
          5 µL
        
        
          Detection:
        
        
          228 nm
        
        
          Mobile Phase:
        
        
          90/10 acetonitrile/100 mM ammonium
        
        
          acetate
        
        
          Flow Rate:
        
        
          1.0 mL/min
        
        
          Inj.Volume:
        
        
          5 µL
        
        
          Detection:
        
        
          248 nm