38
        
        
          80 Å
        
        
          
            Hydrophobicity
          
        
        
          
            pH Range
          
        
        
          
            Pore Size
          
        
        
          2.6 µm
        
        
          14
        
        
          High
        
        
          2 to 8
        
        
          Low
        
        
          0
        
        
          0
        
        
          300
        
        
          
            Particle Size
          
        
        
          The urea phase bonded onto solid core particles provides a unique selectivity for polar analytes.
        
        
          Accucore Urea-HILIC provides good retention for charged and neutral polar compounds.
        
        
          Accucore Urea-HILIC
        
        
          (Tb/Tp)
        
        
          (2dG/3dG)
        
        
          k U /2
        
        
          (CX) /2
        
        
          (OH)
        
        
          (V/A)
        
        
          (AX)
        
        
          (CH2)
        
        
          This neutral material offers good hydrophilic selectivity and
        
        
          low ion exchange activity.
        
        
          
            NH
          
        
        
          
            2
          
        
        
          
            N
          
        
        
          
            H
          
        
        
          
            O
          
        
        
          0.2
        
        
          0.4
        
        
          0.6
        
        
          0.8
        
        
          1.0
        
        
          1.2
        
        
          1.4
        
        
          0
        
        
          200
        
        
          400
        
        
          600
        
        
          1
        
        
          2
        
        
          3
        
        
          Time (min)
        
        
          mAU
        
        
          0
        
        
          Column:
        
        
          Accucore Urea-HILIC
        
        
          , 2.6 µm
        
        
          100 × 2.1 mm
        
        
          Part Number:
        
        
          27726-102130
        
        
          Mobile Phase:
        
        
          composition 10:80:10,  A : B : C
        
        
          A: water
        
        
          B: acetonitrile
        
        
          C: 100 mM ammonium acetate adjusted to pH 4.9
        
        
          Flow Rate:
        
        
          300 µL/min
        
        
          Inj.Volume:
        
        
          2 µL into 10 µL partial loop mode
        
        
          Injection Wash
        
        
          Solvent:
        
        
          water:acetonitrile 20:80
        
        
          Temp.:
        
        
          35 °C
        
        
          Detection:
        
        
          UV at 230 nm
        
        
          Backpressure:
        
        
          71 bar
        
        
          Run Time :
        
        
          2 minutes
        
        
          Acetaminophen
        
        
          Salicylic acid
        
        
          Aspirin
        
        
          t
        
        
          R
        
        
          A
        
        
          s
        
        
          t
        
        
          R
        
        
          A
        
        
          s
        
        
          R
        
        
          s
        
        
          t
        
        
          R
        
        
          A
        
        
          s
        
        
          R
        
        
          s
        
        
          Mean 0.760 1.474 0.908 1.303 2.359 1.100 1.318 3.264
        
        
          CV % 0.00 1.17 0.48 0.92 0.49 0.00 0.63 0.48
        
        
          Data from eight replicate analyses of a mixture of acetaminophen, salicylic acid
        
        
          and aspirin
        
        
          Retention time (t
        
        
          R
        
        
          ), peak asymmetry (A
        
        
          s
        
        
          ), peak resolution (R
        
        
          s
        
        
          )
        
        
          Analgesic compounds