          Column Care in HILIC
        
        
          pH Stability
        
        
          Thermo Scientific HILIC columns are silica based. A suitable pH range is
        
        
          2−8.
        
        
          Washing Conditions
        
        
          Prior to storage, buffered solutions used during HILIC analysis should be
        
        
          washed out of the column (and the HPLC system). A suitable wash
        
        
          solution is 60% organic solvent in water.
        
        
          Storing Conditions
        
        
          HILIC columns should be stored under HILIC conditions. A mixture of
        
        
          organic solvent and water in a 90:10 ratio is recommended for storage.
        
        
          The Product Manual will give specific advice.
        
        
          Ensure the column end-fittings are sealed with end-plugs to prevent the
        
        
          stationary phase from drying.
        
        
          As with any other column, to maximize HILIC column life ensure that samples and mobile
        
        
          phases are particle-free. We also recommend the use of guard columns or inline filters
        
        
          between the switching valve and the column.