The following van Deemter graphs – where plate height, H, is plotted versus linear flow velocity - report the results of
        
        
          this investigation.
        
        
          With respect to the two solid core 2.6 μm particle size materials (Accucore150-Amide-HILIC and Accucore HILIC), it
        
        
          can be seen above that the bare silica has a flatter profile (lower value of the C term in the van Deemter equation) and
        
        
          hence a better preservation of high efficiency at high flow rates than the amide material. The optimum flow rate for
        
        
          the bare silica (Accucore HILIC) is in the range of 0.9−1.7 mL/min; however, even above 1.7 mL/min efficiency is not
        
        
          compromised as the increase in H is minimal.
        
        
          The amide phase (Accucore150-Amide-HILIC) demonstrated an optimum flow rate range of 0.7−1.0 mL/min.
        
        
          Within its optimal flow rate range, the bare silica phase demonstrated slightly lower H values (plate height) than the
        
        
          amide material. Lower H values correspond to higher plate counts, making the bare silica material more efficient.
        
        
          The zwitterionic phase (5 μm particle size) investigated showed a prominent variation in efficiency with flow rate, as
        
        
          illustrated in the figure below:
        
        
          18
        
        
          The zwitterionic phase showed relatively high H and mass transfer term values. The minimum plate height is similar
        
        
          to the values reported by McCalley [13] and Hemström and Irgum [1]. According to Hemström and Irgum, the relative
        
        
          thickness of the zwitterionic interactive layer could be the major contributor to the high mass transfer term (although
        
        
          the authors were referring to a polymeric rather than a monomeric coating).
        
        
          From the van Deemter plot reported above, it can be seen that the best performance on the zwitterionic phase is
        
        
          achieved at flow rates between 0.6−0.8 mL/min.
        
        
          Columns:
        
        
          Accucore 150-Amide-HILIC
        
        
          Accucore HILIC
        
        
          2.6 μm 100 × 4.6 mm
        
        
          Mobile Phase:
        
        
          90/10 acetonitrile/100 mM
        
        
          ammonium formate (pH 3.5)
        
        
          Sample:
        
        
          atenolol
        
        
          Column:
        
        
          Syncronis HILIC
        
        
          , 5 μm
        
        
          100 × 4.6 mm
        
        
          Mobile Phase:
        
        
          90/10 acetonitrile/100 mM ammonium
        
        
          formate (pH 3.5)
        
        
          Sample:
        
        
          atenolol
        
        
          H ( m)
        
        
          Linear velocity (mm/s)
        
        
          2.0
        
        
          3.0
        
        
          4.0
        
        
          5.0
        
        
          6.0
        
        
          7.0
        
        
          8.0
        
        
          9.0
        
        
          10.0
        
        
          11.0
        
        
          12.0
        
        
          0.5
        
        
          0.0
        
        
          1.0
        
        
          1.5
        
        
          2.0
        
        
          2.5
        
        
          3.0
        
        
          3.5
        
        
          4.0
        
        
          Accucore HILIC Accucore 150-Amide-HILIC
        
        
          H ( m)
        
        
          Linear velocity (mm/s)
        
        
          10.0
        
        
          12.0
        
        
          14.0
        
        
          16.0
        
        
          18.0
        
        
          20.0
        
        
          22.0
        
        
          24.0
        
        
          26.0
        
        
          0.5
        
        
          1.0
        
        
          1.5
        
        
          2.0
        
        
          2.5
        
        
          3.0
        
        
          0.0
        
        
          3.5
        
        
          Syncronis HILIC
        
        
          Optimum
        
        
          Flow Rate:
        
        
          Accucore 150-Amide-HILIC
        
        
          , 0.7−1.0 mL/min
        
        
          Accucore HILIC
        
        
          , 0.9−1.7 mL/min
        
        
          Optimum
        
        
          Flow Rate:
        
        
          Syncronis HILIC
        
        
          ,
        
        
          0.6−0.8 mL/min