          2
        
        
          Consumables
        
        
          Part Number
        
        
          Column:
        
        
          TraceGOLD TG-5MS 30 m × 0.25 mm × 0.25 µm 26098-1420
        
        
          Septum:
        
        
          BTO coated 11 mm center guide (50/pk)
        
        
          31303233
        
        
          Liner:
        
        
          Splitless liner with single taper 78.5 × 4 × 6.3 mm 453A1925
        
        
          Column ferrules:
        
        
          Graphite ferrule for 0.1–0.32 mm i.d. columns 10/pk
        
        
          290GA139
        
        
          Injection syringe:
        
        
          10 µL syringe FN 50 mm T Gauge 26, cone tip
        
        
          36500525
        
        
          Vials and closures:
        
        
          Thermo Scientific 9 mm Wide Opening Screw Thread 60180-599
        
        
          Vials Convenience Kit, 2 mL Clear Glass Vial with ID Patch,
        
        
          Blue Closure with PTFE/Blue Silicone Septa
        
        
          Solutions
        
        
          A working standard solution of 20 µg/mL of EPA 8141 was prepared in acetone. The stock solution was obtained
        
        
          commercially at a concentration of 1000 µg/mL.
        
        
          Separation Conditions
        
        
          Part Number
        
        
          Instrumentation:
        
        
          TRACE 1310 mainframe 230 V GC
        
        
          14800302
        
        
          Carrier gas:
        
        
          Helium
        
        
          Split flow:
        
        
          50 mL/min
        
        
          Column flow:
        
        
          1.2 mL/min, constant flow
        
        
          Oven temperature:
        
        
          40 °C (1 min), 12 °C/min, 280 °C (10 min)
        
        
          Injector type:
        
        
          TRACE 1310 SSL Injector module
        
        
          29903010
        
        
          Injector mode:
        
        
          Splitless
        
        
          Injection details:
        
        
          Splitless (1 min)
        
        
          Injector temperature:
        
        
          220 °C
        
        
          Detector details:
        
        
          TRACE 1310 FID module                                                 29903001
        
        
          FID parameters:
        
        
          Temperature:
        
        
          280 °C
        
        
          Air flow:
        
        
          350 mL/min
        
        
          Hydrogen flow:
        
        
          35 mL/min
        
        
          Nitrogen makeup flow:
        
        
          30 mL/min
        
        
          Injector Conditions
        
        
          Instrumentation:
        
        
          Thermo Scientific AS1300 Autosampler
        
        
          Injection Volume:
        
        
          1 µL
        
        
          Wash solvent:
        
        
          Acetone/hexane (1:1 v/v)
        
        
          Data Processing
        
        
          Software:
        
        
          Thermo Scientific
        
        
          TM
        
        
          Chrom-Card
        
        
          TM
        
        
          data system
        
        
          Results
        
        
          Figure 1 shows the TIC chromatogram for 22 OPPs at 20 µg/µL using a TraceGOLD TG-5MS
        
        
          column and a standard, deactivated, splitless quartz liner for the TRACE 1310 GC instrument.
        
        
          Table 1 shows the peak identification of the OPPs according to their retention times. Table 1
        
        
          includes the reproducibility data for ten injections. The stationary phase of the TG-5MS GC
        
        
          column, in combination with the deactivated splitless liner, provides excellent performance due to
        
        
          minimal interaction of active compounds with active sites on the column, the glass wall of the liner,
        
        
          or the deactivated quartz wool. This minimizes peak tailing of the OPPs and gives  highly
        
        
          symmetrical peak shapes. The combination of a TG-5MS GC column, the deactivated liner, and the
        
        
          TRACE 1310 GC gave excellent injection reproducibility of between 1.7% and 3.4% for the 22
        
        
          OPPs tested (Table 2).