 
          2
        
        
          Experimental Details
        
        
          Consumables
        
        
          Part Number
        
        
          Columns:
        
        
          TraceGOLD TG-5SilMS, 30 m x 0.25 mm x 0.25 μm 26096-1420
        
        
          GuardGOLD 2 m x 0.53 mm ID
        
        
          26050-0253
        
        
          Press-Fit
        
        
          64000-001
        
        
          Injection port septum:
        
        
          BTO, 12.7 mm
        
        
          313032280
        
        
          Liner:
        
        
          PTV Silcosteel liner for PTV simulated on-column,
        
        
          45322052
        
        
          1 x 2.75 x 120 mm
        
        
          Column ferrules:
        
        
          100% graphite ferrules for Thermo Scientific
        
        
          29053486
        
        
          TRACE™ injector, 0.53 mm ID
        
        
          Colum ferrules:
        
        
          Graphite/Vespel
        
        
          ®
        
        
          for transfer line, 0.25 mm ID
        
        
          29033496
        
        
          Injection syringe:
        
        
          85 mm 26s Gauge, 10 μL fixed needle syringe
        
        
          365D0321
        
        
          for a Thermo Scientific TriPlus™ RSH Autosampler
        
        
          Sample vials:
        
        
          Thermo Scientific National™ 9 mm
        
        
          C4000-11
        
        
          Target DP Polypropylene Vial, 300 µL
        
        
          Thermo Scientific National 9 mm Target DP™ Vial,
        
        
          C4000-9TR
        
        
          Total Recovery with 10 µL Reservoir
        
        
          Thermo Scientific Chromacol™ 9 mm screw caps 9-SC(B)-ST101
        
        
          with High Purity Silicone/PTFE septa
        
        
          LC-MS grade water
        
        
          Preparation of Calibration Standards
        
        
          A stock standard solution of 1 mg/mL of bifenthrin, permethrin, cyfluthrin, and cypermethrin was prepared in ethyl
        
        
          acetate. Calibration standard solutions were then prepared in ethyl acetate at the following concentrations: 50, 100, 200,
        
        
          500, 1000, and 2000 ng/mL. A 100 µL aliquot of each calibration standard was then placed into an autosampler vial
        
        
          followed by the addition of 10 µL of 10 µg/mL of internal standard to each vial.
        
        
          Sample Preparation: SPE Extraction Protocol
        
        
          Part Number
        
        
          SPE cartridge:
        
        
          Thermo Scientific HyperSep™ C18 SPE column,
        
        
          60108-701
        
        
          2000 mg/15 mL
        
        
          Compound:
        
        
          (i) Bifenthrin, (ii)
        
        
          cis/trans
        
        
          permethrin, (iii) cyfluthrin,
        
        
          and (iv) cypermethrin 1 L each at 0.10 ng/mL in water
        
        
          Matrix:
        
        
          LC/MS water
        
        
          Conditioning stage:
        
        
          10 mL ethyl acetate, 10 mL acetone, 2 x 10 mL aliquots water applied
        
        
          sequentially to the SPE cartridge and then pulled through under
        
        
          vacuum at 4-5 mL/min
        
        
          Application stage:
        
        
          1 L of sample was applied to the SPE cartridge under vacuum
        
        
          at 4-5 mL/min
        
        
          Washing stage:
        
        
          10 mL of water was added to a 1 L vessel, swirled, and placed onto the
        
        
          SPE cartridge. Then the cartridge was dried for 20 min under vacuum.
        
        
          Elution stage:
        
        
          10 mL ethyl acetate was added to the sample vessel, swirled, and then
        
        
          placed onto the SPE cartridge. Then an additional 10 mL ethyl acetate
        
        
          was applied directly onto the cartridge.
        
        
          Additional stages:
        
        
          Solvent was evaporated to dryness at 40 °C and the residue
        
        
          reconstituted in 100 µL of ethyl acetate to give the final concentration
        
        
          of pyrethroids at 1000 ng/mL. Then 10 μL of 10 μg/mL of internal
        
        
          standard was added to the vial.