 
          
            Experimental Conditions
          
        
        
          All samples were prepared using the QuEChERS technique,
        
        
          and calibration was performed using a blank QuEChERS
        
        
          extract from cucumber. All target compounds were
        
        
          measured using at least two SRM transitions for each
        
        
          compound to a level of 0.001 mg/kg, which is ten times
        
        
          lower than the current maximum concentration limit.
        
        
          All sample analyses were carried out using the TSQ
        
        
          Quantum XLS Ultra GC-MS/MS system, equipped with
        
        
          a Thermo Scientific TRACE GC Ultra gas chromatograph.
        
        
          The TRACE GC Ultra
        
        
          ™
        
        
          was configured with a B.E.S.T.
        
        
          PTV injector equipped with a backflush device. Sample
        
        
          introduction was performed using the Thermo Scientific
        
        
          TriPlus RSH autosampler. The capillary column was a
        
        
          Thermo Scientific TraceGOLD TG-5MS column (5%
        
        
          phenyl film) of 30 m length, 0.25 mm inner diameter
        
        
          and 0.25 μm film thickness (Table 1).
        
        
          The pre-column used was a 1.2 m TG-5HT, 0.15 μm
        
        
          film thickness and 0.53 mm inner diameter (see Table 1).
        
        
          
            Maximizing Robustness
          
        
        
          High boiling compounds in sample matrix have a negative
        
        
          effect on the analytical column’s quality and lifetime,
        
        
          requiring a bake out process at high temperatures, thus
        
        
          limiting sample throughput. A backflush process was
        
        
          used to protect the column, allowing more samples to be
        
        
          injected before the phase attachment on the surface of
        
        
          the column becomes weak. Being able to inject more
        
        
          samples before necessary column replacement improves
        
        
          throughput and reduces costs per analyses.
        
        
          During backflushing of the pre-column, the injector was
        
        
          set to a higher temperature and increased flow. This also
        
        
          allowed the injector liner to be swept of residual matrix
        
        
          contaminants during analysis time. This concurrent
        
        
          backflush operation results in the complete system staying
        
        
          clean and inert for a high number of injections, resulting
        
        
          in less maintenance frequencies.
        
        
          3
        
        
          
            Method Setup
          
        
        
          The method parameters for the PTV concurrent backflush
        
        
          operation, GC separation and TSQ Quantum XLS Ultra
        
        
          mass spectrometer setup are given in Table 1.
        
        
          Each compound SRM transition was only monitored for
        
        
          a narrow time window around the established retention
        
        
          time (timed SRM). This led to a fully optimized instrument
        
        
          duty cycle for maximum analytical performance being
        
        
          handled automatically by the system. The complete list
        
        
          can be copied into the instrument method, thus saving
        
        
          time and avoiding entry errors.
        
        
          4
        
        
          For data acquisition, the two most selective transitions were
        
        
          chosen after reviewing data from spiked matrix samples.
        
        
          Selection criteria were based on the absence of interferences
        
        
          from the matrix, along with signal generation of the
        
        
          transition.
        
        
          
            Results and Discussion
          
        
        
          
            Advanced GC-MS/MS Experiments – U-SRM
          
        
        
          The patented Thermo Scientific HyperQuad technology
        
        
          in the TSQ Quantum XLS Ultra system offered high
        
        
          sensitivity by high ion transmission already found at
        
        
          the standard nominal mass resolution settings (0.7 Da
        
        
          FWHM). In addition, the HyperQuad
        
        
          ™
        
        
          technology allows
        
        
          the possibility to enhance the applied mass resolution for
        
        
          increased selectivity during analysis. The significantly
        
        
          increased selectivity further reduces the background caused
        
        
          by matrix components, thus giving a cleaner peak detection
        
        
          and high signal-to-noise results.
        
        
          Some compound transitions are more susceptible to
        
        
          matrix interference than others. Standard SRM resolution
        
        
          (0.7 Da) can often provide enough selectivity to overcome
        
        
          most matrix interference challenges. In complex matrices,
        
        
          however, even with the structure-selective SRM acquisitions,
        
        
          removal of the isobaric matrix interference is insufficient.
        
        
          2
        
        
          
            TRACE GC Ultra
          
        
        
          Injection Volume
        
        
          2 μL injection
        
        
          Liner
        
        
          Siltec
        
        
          ®
        
        
          baffled liner
        
        
          (part number 453T2120)
        
        
          Carrier Gas
        
        
          He, constant flow, 1.3 mL/min
        
        
          Column Type
        
        
          TraceGOLD
        
        
          ™
        
        
          TG-5MS column (5% phenyl film) of
        
        
          30 m length, 0.25 mm inner diameter and 0.25 μm
        
        
          film thickness (part number 26098-1420)
        
        
          Precolumn
        
        
          1.2 m of TraceGOLD TG-5HT column of 30 m
        
        
          length, 0.53 mm inner diameter and 0.15 µm film
        
        
          thickness (part number 26095-0620)
        
        
          GC Method
        
        
          Initial 65 °C, Hold 1.5 min,
        
        
          Ramp 30.0 °C/min–150 °C,
        
        
          Ramp 5.0 °C/min–290 °C,
        
        
          Ramp 30.0 °C/min–320 °C, Hold 5.0 min
        
        
          Transfer Line
        
        
          300 °C
        
        
          
            TRACE GC Ultra PTV Program
          
        
        
          Injector Temperature
        
        
          70 °C, splitless injection 1.5 min
        
        
          PTV Inject
        
        
          70 °C, 0.2 min, 8 °C/sec to transfer step
        
        
          PTV Transfer
        
        
          280 °C, 21 min, 10 °C/sec to clean step
        
        
          PTV Clean
        
        
          350 °C, 33 min, clean flow 30 mL/min
        
        
          Transfer Time
        
        
          21 min
        
        
          
            TSQ Quantum XLS Ultra Mass Spectrometer
          
        
        
          Source Temperature
        
        
          240 °C, CEI volume
        
        
          Ionization
        
        
          EI, 70 eV
        
        
          Emission Current
        
        
          50 μA
        
        
          Resolution
        
        
          0.7 Da Q1, Q3; 0.1 Da on Q1, 0.7 on Q3 for the
        
        
          wheat examples
        
        
          Collision Gas
        
        
          Argon, 1.5 mTorr
        
        
          Table 1: Selected instrument conditions for the employed TRACE GC
        
        
          Ultra and TSQ Quantum XLS Ultra mass spectrometer