AI10382-GC-MS-Food Safety-Analysis - page 37

Application Note 52279
As an additional test, the ion ratio at different levels has
been monitored and the deviation of the transitions has
been calculated.
Compound
Ion Ratio Deviation RSD in % (n=5)
Phosmet
0.79
Azinphosmethyl
3.65
Chlorfenapyr
15.08
Deltamethrin
0.88
Iprodion
5.34
Alfa Endosulfan
3.63
Methidathion
0.84
Carbaryl
3.64
Cyfluthrin
3.55
Pyrimifos
3.83
Table 3: Ion ratio deviation of some challenging pesticides in
cucumber matrix at several levels of concentration
Figure 3 (a) through (f) show a 1 ppb matrix spike and
calibration data obtained for select targeted pesticides in
cucumber matrix.
Conclusions
Advances in HyperQuad technology offers increased
analytical performance for routine applications such as
pesticide analysis.
A true multi-compound method was developed for over
400 pesticides using timed SRM; easily transferable
from a spreadsheet.
A high level of accuracy and precision was reached
during data evaluation, on several cornerstones of
analysis, such as repeatability, linearity and ion ratio
stability.
Furthermore, all examples shown are the more
challenging pesticides faced analytically in terms of
stability, activity and response.
This resolution technology development allows for
advanced GC-MS/MS operations to be performed,
such as U-SRM to further increase selectivity in
complex matrices. This not only improves quantitative
measurements, but it is also amenable when using a
reduced sample clean-up which is typical for QuEChERS
methodologies.
References
1. REGULATION (EC) No 396/2005 and amendments, Feb. 23 2005.
2. M. Anastassiades, S. Lehotay D. Steinbacher, F. Schenck,
J. AOAC
Int.
86 (2) (2003) 412.
3. Munari, F., Huebschmann, H.J., Eliminating High Boiling Matrix in
GC and GC/MS by Using PTV Backflush Injection Technique for
Increased Productivity and Reliability, Application Note 51888,
Thermo Fisher Scientific, 2010.
4. Pesticides Method Reference, 2nd ed. 2011, Thermo Fisher Scientific,
Austin, TX, USA, (part number 120390).
For reference on Thermo Scientific QuEChERS products, please see our catalog, Thermo
Scientific HyperSep Dispersive SPE Products (part number BRGSCQUECHERS 1109).
AN52279_E 08/12M
Africa-Other
+27 11 570 1840
Australia
+61 3 9757 4300
Austria
+43 1 333 50 34 0
Belgium
+32 53 73 42 41
Canada
+1 800 530 8447
China
+86 10 8419 3588
Denmark
+45 70 23 62 60
Europe-Other
+43 1 333 50 34 0
Finland/Norway/Sweden
+46 8 556 468 00
France
+33 1 60 92 48 00
Germany
+49 6103 408 1014
India
+91 22 6742 9434
Italy
+39 02 950 591
Japan
+81 45 453 9100
Latin America
+1 561 688 8700
Middle East
+43 1 333 50 34 0
Netherlands
+31 76 579 55 55
New Zealand
+64 9 980 6700
Russia/CIS
+43 1 333 50 34 0
South Africa
+27 11 570 1840
Spain
+34 914 845 965
Switzerland
+41 61 716 77 00
UK
+44 1442 233555
USA
+1 800 532 4752
©2012 Thermo Fisher Scientific Inc. All rights reserved. ISO is a trademark of the International Standards Organization.
All other trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. This information is presented as an
example of the capabilities of Thermo Fisher Scientific Inc. products. It is not intended to encourage use of these products in any
manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change.
Not all products are available in all countries. Please consult your local sales representative for details.
Thermo Fisher Scientific,
Austin, TX USA is ISO Certified.
1...,27,28,29,30,31,32,33,34,35,36 38,39,40,41,42,43,44,45,46,47,...225
Powered by FlippingBook