AI10382-GC-MS-Food Safety-Analysis - page 210-211

3
2
This technical note describes results obtained from an
endurance mechanical test performed on “Instant Connect”
modules and typical analytical reproducibility and
accuracy provided by the TRACE 1300 Series GC.
Analytical reproducibility was also measured after switching
modules of the same type in the same application, as
might happen when an injector module is replaced with a
new one in a routine laboratory to keep a contaminated
instrument up and running.
Experimental
A TRACE 1310 GC instrument equipped with various
SSL and FID modules was used in all of the experiments,
and all “Instant Connect” modules are identified by specific
serial numbers for easier tracking in the lab. Four different
modules were alternated in these experiments.
The GC was equipped with a Thermo Scientific AS 1310
liquid autosampler. All tests were performed using a
synthetic mix of normal alkanes ranging from C
10
to C
40
in hexane, at a concentration of approximately 10 ppm
(10 ng/μL) and using helium as carrier gas. A 1 μL aliquot
of sample was injected in splitless mode into a standard
glass-wool packed tapered liner, while the injector
temperature was maintained at 300 °C. Splitless time
was 0.8 minutes. The FID detector temperature was set
to 350 °C.
A Thermo Scientific TraceGOLD
column TR-5, 15 m x
0.25 mm id x 0.25 μm, was used in all experiments. Oven
temperature was set at 50 °C for 0.5 min and then ramped
up to 340 °C at 20 °C/min, with two minutes of isothermal
time at the final temperature. A Thermo Scientific Dionex
Chromeleon
Chromatography Data System was used for
setting all method parameters, data acquisition, and data
processing.
Results and Discussion
System analytical reproducibility and accuracy
The analytical reproducibility was evaluated using two
new “Instant Connect” modules, a SSL injector
(module serial number S/N: 712100036) and a FID
detector (S/N: 712300088), by injecting the synthetic
hydrocarbon mix automatically (ten repetitions). Results in
terms of peak area and retention time repeatability are
summarized in Tables 1 and 2. No discrimination for both
volatile and high boiling compounds was seen. As shown
in Table 1 the recovery, calculated using C
20
peak area as
reference, is close to 100% along the full range of
volatility. Absolute peak area relative standard deviation
is far below 1% for all hydrocarbons. All injector and
detector modules incorporate a new generation of
miniaturized gas controls. These integrated electronic
devices ensure precise control of the inlet pressure and the
flow throughout the column, further contributing to the
excellent reproducibility of retention times. As indicated
in Table 2, the standard deviation is below a thousandth of
a minute. This level of reproducibility is a clear indication
of the accurate temperature profile and column flow
maintained during the ramp and the precise thermo-
regulation of the GC oven. Overall results show full
recovery of hydrocarbons and excellent data precision.
Table 1. Instrument configuration SSL S/N 712100036 and FID S/N 712300088. Absolute peak area RSD% far lower than 1%. Recovery,
measured as ratio vs C
20
average area, at 100% for the whole range of hydrocarbons
SSL 712100036
/ FID 712300088
nC10
nC12
nC14
nC16
nC18
nC20
nC22
nC24
nC26
nC28
nC30
nC32
nC34
nC36
nC38
nC40
Injection 1
2600304
2647767
2600941
2619188
2552750
2565809
2543886
2535687
2512699
2527008
2602759
2597807
2534441
2564855
2470270
2525384
Injection 2
2610605
2657856
2601653
2623404
2568557
2579380
2565938
2565679
2545232
2560614
2636265
2629734
2557462
2596729
2495209
2563483
Injection 3
2602666
2653832
2599714
2626029
2553641
2577265
2561577
2558672
2542703
2555406
2632496
2630095
2555743
2580475
2480864
2538819
Injection 4
2636572
2683702
2632897
2656448
2593709
2602685
2584957
2575384
2558834
2574920
2649582
2640082
2566623
2593858
2490053
2549873
Injection 5
2623737
2668679
2617130
2639475
2575209
2588255
2568857
2566489
2551218
2570336
2641759
2632243
2559472
2591033
2487269
2545848
Injection 6
2628675
2671731
2625320
2647746
2586155
2602674
2584119
2578956
2563433
2577945
2652932
2644762
2572449
2600568
2495549
2560371
Injection 7
2633245
2675436
2621623
2640507
2579749
2601553
2603546
2589030
2566470
2580193
2651340
2644782
2575086
2615870
2515378
2552861
Injection 8
2622426
2667773
2618401
2631007
2571368
2588047
2571982
2568771
2543992
2565899
2635937
2628421
2556233
2599156
2491820
2552234
Injection 9
2627383
2675413
2624978
2646945
2578061
2590137
2578171
2582555
2553973
2565982
2637795
2636002
2561603
2598494
2504800
2575965
Injection 10
2621650
2664829
2611668
2634863
2576839
2592681
2577082
2571091
2552087
2567396
2634338
2631176
2560023
2590260
2497076
2558360
Average (counts)
2620726
2666702
2615432
2636561
2573604
2588848
2574011
2569231
2549064
2564570
2637520
2631510
2559913
2593130
2492829
2552320
SD
12355
10941
11571
11889
12894
12092
15913
14767
15078
15223
14298
13319
11161
13430
12358
13908
RSD %
0.47%
0.41%
0.44%
0.45%
0.50%
0.47%
0.62%
0.57%
0.59%
0.59%
0.54%
0.51%
0.44%
0.52%
0.50%
0.54%
Recovery %
101%
103%
101%
102%
99%
100%
99%
99%
98%
99%
102%
102%
99%
100%
96%
99%
Figure 1. Module-to-module repeatability. Modules store all
of their calibration information allowing minimum variation if
replaced on a system.
Module-to-module reproducibility
To simulate a situation where a laboratory needs to quickly
replace a module, such as to avoid interrupting instrument
throughput for maintenance, the “Instant Connect” SSL
injector module (S/N: 712100036) was replaced by a new
module (S/N: 712300021). This required cooling and
powering down the instrument, disconnecting the column
from the original SSL injector module, removing the module
and plugging in the new one, connecting the column, and
powering up the TRACE 1310 GC again. Electronic gas
control permits an automated leak check to be performed
to guarantee that no artifacts are introduced by this
manual operation. The reduced thermal mass GC design
allows a quick recovery of injection-ready conditions after
instrument power-up. As a result, the GC was ready to
resume analytical injections again in only nine minutes
after it was originally powered down. A blank GC cycle
was programmed before injecting samples again, which is
good practice to ensure the entire flow path was not
affected by air introduced during module replacement.
An automated sequence of 10 injections was performed
immediately after the module replacement along with
collecting data. The instrument was then stopped again,
and the FID detector module (S/N: 712300088) was
replaced by a new one (S/N: 712300126). After a blank
run, another sequence of 10 injections completed the
experiments. Tables 3 and 4 and Figure 1 summarize the
repeatability results for the three different instrument
configurations. Variations in peak area measured as a
delta of the average counts are in the range of a few
percentages when changing either the injector or the FID
detector. Such a variation, for many applications, is well
below the required limit of a system suitability check,
eliminating the need to recalibrate the GC system as a
whole. The retention time variations are in the range of a
few hundredths of a minute or even less with no impact
on component retention time.
1...,200,201,202,203,204,205,206,207,208,209 212-213,214-215,216,217,218,219,220,221,222,223,...225
Powered by FlippingBook