2
        
        
          interferences in the analysis. The same effect can be
        
        
          observed when mass resolution settings of 1.2 Da FWHM
        
        
          or higher are used during single quadrupole analysis. This
        
        
          can be seemingly effective when simple sample matrices
        
        
          are encountered. However, when facing a more complex
        
        
          sample matrix this strategy can cause problems in the
        
        
          analysis.
        
        
          What are Isobaric Interferences?
        
        
          The term “isobaric interferences” describes ions of the
        
        
          same nominal mass but of different chemical composition
        
        
          and structure. The “nominal mass” is typically the mass
        
        
          used in quadrupole instruments for programming SIM or
        
        
          SRM acquisition. This reflects the typical unit mass
        
        
          resolution capabilities of standard quadrupole instru-
        
        
          ments. The nominal mass, in this context, can be
        
        
          described as a 1 Da (1
        
        
          
            m/z
          
        
        
          ) resolution capability between
        
        
          mass peaks. The term “isobaric interference” means in
        
        
          practice that ion signals from other compounds than the
        
        
          target analyte appear at the same nominal mass in the
        
        
          scan spectrum, the SIM trace or, in some cases, the SRM
        
        
          trace.
        
        
          For triple quadrupole instruments operated in SRM
        
        
          mode, the selectivity is generally high. This is due to the
        
        
          MS/MS process. There are cases, however, where despite
        
        
          MS/MS being applied, selectivity is challenged. These
        
        
          cases appear more frequently when the matrix load of
        
        
          samples is very high. This is not unusual for a triple
        
        
          quadrupole, as often the most complex quantitative
        
        
          determinations are directed to this type of technology. The
        
        
          likelihood of encountering a full SRM interference
        
        
          increases as a function of the matrix complexity. The
        
        
          observed effects of isobaric interferences are also more
        
        
          apparent when targeting compounds in low or sub-ppb
        
        
          concentration ranges. This is because the target com-
        
        
          pound mass is more likely to be “shadowed” by
        
        
          interfering matrix ions (especially in the first stage of MS
        
        
          in Q1) that are typically orders of magnitude higher in
        
        
          concentration. The drive towards shorter clean-up
        
        
          procedures also pushes additional matrix to the detection
        
        
          system, adding to the problem.
        
        
          When these intense interfering matrix ions successfully
        
        
          transmit through Q1 into the collision cell, there is a
        
        
          higher statistical probability that interfering product ion
        
        
          masses are formed. This gives rise to a higher occurrence
        
        
          of full SRM interferences and visibly reduced analyte
        
        
          selectivity. This often manifests as an increased chemical
        
        
          noise background and hence, low signal-to-noise detec-
        
        
          tion. This is observed most frequently in matrix samples
        
        
          and is often unnoticed or absent in solvent only standards
        
        
          (see Figure 3). Because solvent standards are relatively
        
        
          clean, it is also possible (and sometimes practiced) to open
        
        
          Q1 above unit mass resolution (so called “wide” or
        
        
          “open” settings.) This creates the possibility to gain
        
        
          sensitivity, but does not help much when considering real
        
        
          backgrounds in complex samples. Sensitivity (and
        
        
          selectivity) achieved at wider Q1 resolution values can
        
        
          vary considerably between clean and dirty samples. With
        
        
          that in mind, it is sensible that any comparisons between
        
        
          instruments, especially those that are to face dirtier
        
        
          samples, are performed in matrix samples. If solvent
        
        
          standards are to be used, then the true instrument
        
        
          sensitivity should be compared using equivalent Q1 and
        
        
          Q3 resolution values.
        
        
          How Triple Quadrupole Analyzers Work
        
        
          The idea of using three quadrupoles arranged in series in a
        
        
          triple quadrupole analyzer for structure elucidation (“an
        
        
          added dimension of mass spectral information”) follows
        
        
          an idea first reported by Richard Yost and Chris Enke of
        
        
          Florida University in 1970. The analyzer should allow the
        
        
          detection of structure-related information and overcome
        
        
          the single quadrupole limitation of measuring a mass
        
        
          The observed effects of isobaric
        
        
          interferences are also more apparent when
        
        
          targeting compounds in low or sub-ppb
        
        
          concentration ranges.
        
        
          Figure 2. HyperQuad quadrupole rods used in the TSQ Quantum
        
        
          XLS Ultra system
        
        
          Figure 3. High occurrence of matrix interference on standard
        
        
          specification compound (octafluoronapthalene) with “Wide/
        
        
          Open” Q1 settings (1.2 Da FWHM) absent in solvent standard
        
        
          (left) but significant when in presence of complex matrix (right).
        
        
          1.2 Da wide resolution
        
        
          100 fg/µL in
        
        
          clean solvent
        
        
          1.2 Da wide resolution
        
        
          100 fg/µL in diesel
        
        
          matrix