Thermo Fisher Scientific,
Austin, TX USA is ISO Certified.
TN52364_E 11/12S
Africa-Other
+27 11 570 1840
Australia
+61 3 9757 4300
Austria
+43 1 333 50 34 0
Belgium
+32 53 73 42 41
Canada
+1 800 530 8447
China
+86 10 8419 3588
Denmark
+45 70 23 62 60
Europe-Other
+43 1 333 50 34 0
Finland/Norway/Sweden
+46 8 556 468 00
France
+33 1 60 92 48 00
Germany
+49 6103 408 1014
India
+91 22 6742 9434
Italy
+39 02 950 591
Japan
+81 45 453 9100
Latin America
+1 561 688 8700
Middle East
+43 1 333 50 34 0
Netherlands
+31 76 579 55 55
New Zealand
+64 9 980 6700
Russia/CIS
+43 1 333 50 34 0
South Africa
+27 11 570 1840
Spain
+34 914 845 965
Switzerland
+41 61 716 77 00
UK
+44 1442 233555
USA
+1 800 532 4752
©2012 Thermo Fisher Scientific Inc. All rights reserved. ISO is a trademark of the International Standards Organization.
All other trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. This information is presented as an
example of the capabilities of Thermo Fisher Scientific Inc. products. It is not intended to encourage use of these products in any
manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change.
Not all products are available in all countries. Please consult your local sales representative for details.
Technical Note 52364
Conclusion
The “Instant Connect” modules on the TRACE 1300
Series GC offer important advantages over conventional
GC instrumentation, such as maintaining instrument
uptime and continuing to run even when an injector or
detector must be replaced for maintenance purposes.
Additional advantages include the ability to upgrade from
a single-channel GC to a double-channel GC to increase
instrument productivity and the option to add a new
detector module to respond to new application requirements.
The design of “Instant Connect” modules as self-independent
components of the GC, which incorporate all mechanical
and electronic components with calibration information,
permits the user to rapidly remove and install new modules
without any service assistance. The modularity of the
design provides configuration flexibility never before
available in a GC and also maintains the highest
reproducibility and ruggedness standards. The test results
show that module internal calibration allows module-to-
module reproducibility to be within 5% of the variances
in absolute peak area and retention times. The compact
size and robustness of the module design enable the
modules to be repeatedly replaced without impacting
instrument performance.