          2
        
        
          Reagents
        
        
          Part Number
        
        
          Thermo Scientific MBTFA 10 x 1 mL ampules
        
        
          TS-49700
        
        
          Thermo Scientific pyridine silylation grade solvent
        
        
          TS-27530
        
        
          Sample Handling Equipment
        
        
          Part Number
        
        
          Thermo Scientific Reacti-Therm III Heating/Stirring Module
        
        
          TS-18823
        
        
          Thermo Scientific Reacti-Vap III Evaporator
        
        
          TS-18826
        
        
          Thermo Scientific Reacti-Block Q-1 (Holds 8 x 10 mL Reacti-Vials)
        
        
          TS-18814
        
        
          Thermo Scientific Reacti-Vial clear glass reaction vials 10 mL
        
        
          TS-13225
        
        
          Thermo Scientific 2 mL amber vial and screw tops
        
        
          60180-565
        
        
          Separation Conditions
        
        
          Part Number
        
        
          Instrumentation:
        
        
          Thermo Scientific TRACE GC Ultra
        
        
          Column:
        
        
          TRACE™ TR-1701
        
        
          260Q142P
        
        
          30 m × 0.25 mm × 0.25 μm
        
        
          Thermo Scientific BTO 17 mm septa
        
        
          31303211
        
        
          5 mm ID focus split liner, 105 mm long
        
        
          453T1905
        
        
          Graphite liner seal
        
        
          29033406
        
        
          10 μL, 50 mm needle length gauge 25 Syringe
        
        
          36500525
        
        
          Graphite ferrules to fit 0.1-0.25 mm ID columns
        
        
          29053488
        
        
          Carrier gas:
        
        
          Helium
        
        
          Split flow:
        
        
          60 mL/min
        
        
          Column flow:
        
        
          1.2 mL/min, Constant flow
        
        
          Split ratio:
        
        
          1:1
        
        
          Oven temperature:
        
        
          40 °C (1 min), 10 °C/min, 260 °C (5 min)
        
        
          Injector type:
        
        
          Split/Splitless
        
        
          Injector mode:
        
        
          Split
        
        
          Injector temperature:
        
        
          200 °C
        
        
          Detector type:
        
        
          FID
        
        
          Detector temperature:
        
        
          250 °C
        
        
          Detector air flow:
        
        
          35 mL/min
        
        
          Detector Hydrogen flow:
        
        
          350 mL/min
        
        
          Detector nitrogen flow:
        
        
          30 mL/min
        
        
          Thermo Scientific TriPlus Autosampler
        
        
          Injection Volume:
        
        
          1 μL
        
        
          Data Processing
        
        
          Software:
        
        
          Thermo Scientific XCalibur