4. Precision

A) CSS samples: Precision was first assessed with spiked CSS

at two concentration levels (12 and 300 pg/mL). Inter-

(n=15) and intra-batch (n=5) CV values ranged between

1.6% to 12.5% (Table 2).

B) Pooled human plasma samples: Precision was also as-

sessed with a spiked human plasma pool (35.4 pg/mL

of E1 and 18.1 pg/mL of E2, n=5) and the determined

intra-batch CV was 2.2% and 3.6% for E1 and E2,

respectively.

Retention Time (min)

Intensity

2.1

2.4

2.7

3.0

3.3

3.6

3.9

0

82

E1:

E2:

SRM: 269 145

3.5 pg/mL

SRM: 271 183

4.4 pg/mL

Figure 4. SRM chromatograms of E1 and E2 at their LLOQ in spiked CSS

E1

E2

Charcoal Stripped Serum

Low (12 pg/mL)

High (300 pg/mL)

Low (12 pg/mL)

High (300 pg/mL)

Batch 1 Intra-assay Precision (n=5, %)

7.1

6.9

9.4

6.7

Batch 2 Intra-assay Precision (n=5, %)

5.5

1.6

12.5

3.0

Batch 3 Intra-assay Precision (n=5, %)

7.2

4.9

8.0

3.1

Batch 1-3 Inter-assay Precision (n=15, %)

7.3

4.7

10.9

4.4

Spiked Pooled Plasma

E1 (35.4 pg/mL) E2 (18.1 pg/mL)

Precision (n=5, %)

2.2

3.6

Table 2. Precision data