Conclusion

We have developed a novel 10-min LC-MS/MS method

for quantitation of E1 and E2 in serum using TurboFlow

technology for clinical research laboratories. This method

is fast and analytically sensitive and sample preparation

effort is significantly reduced. The Accucore HPLC column

was used for analytical LC separation because of its

superior performance. The lower limit of quantitation was

3.8 pg/mL for estrone and 3.7 pg/mL for estradiol. This

method was linear from 3.8 to 1000.9 pg/mL for estrone

and 3.7 to 993.1 pg/mL for estradiol with accuracy from

95.5% to 103.2% for estrone and from 92.7% to 112.3%

for estradiol, respectively. Inter-assay and intra-assay CV

for estrone and estradiol at low and high concentration

levels in both spiked charcoal stripped serum and pooled

human plasma ranged from 3.5% to 18.0%.

www.thermoscientific.com

Legal Notices: ©2012 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries.

This information is presented as an example of the capabilities of Thermo Fisher Scientific Inc. products. It is not intended to encourage use of these products in any

manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all

countries. Please consult your local sales representative for details.

Thermo Fisher Scientific,

San Jose, CA USA is ISO Certified.

In addition to these

offices, Thermo Fisher

Scientific maintains

a network of represen­

tative organizations

throughout the world.

Africa-Other

+27 11 570 1840

Australia

+61 3 9757 4300

Austria

+43 1 333 50 34 0

Belgium

+32 53 73 42 41

Canada

+1 800 530 8447

China

+86 10 8419 3588

Denmark

+45 70 23 62 60

Europe-Other

+43 1 333 50 34 0

Finland/Norway/

Sweden

+46 8 556 468 00

France

+33 1 60 92 48 00

Germany

+49 6103 408 1014

India

+91 22 6742 9434

Italy

+39 02 950 591

Japan

+81 45 453 9100

Latin America

+1 561 688 8700

Middle East

+43 1 333 50 34 0

Netherlands

+31 76 579 55 55

New Zealand

+64 9 980 6700

Russia/CIS

+43 1 333 50 34 0

South Africa

+27 11 570 1840

Spain

+34 914 845 965

Switzerland

+41 61 716 77 00

UK

+44 1442 233555

USA

+1 800 532 4752

AN63519_E 01/12S

Reference

1.

Xiang He and Marta Kozak, Fast and Sensitive LC-APCI-MS/MS

Quantitative Analysis of Estrone and Estradiol in Serum without Chemical

Derivatization, Thermo Fisher Scientific Application Note 530.