Application Note 603
Table 3. QC and unknowns
Conclusion
A high-throughput, cost-efficient research method was
developed for the precise and accurate measurement of
immunosuppressant drugs in dried blood spots using a
TSQ Endura triple quadrupole mass spectrometer. This
method met analytical laboratory precision and accuracy
criteria for 3 mm dried blood spots.
For research use only. Not for use in diagnostic procedures.
AN64053-EN 0714S
Africa
+43 1 333 50 34 0
Australia
+61 3 9757 4300
Austria
+43 810 282 206
Belgium
+32 53 73 42 41
Canada
+1 800 530 8447
China
800 810 5118
(free call domestic)
400 650 5118
Denmark
+45 70 23 62 60
Europe-Other
+43 1 333 50 34 0
Finland
+358 9 3291 0200
France
+33 1 60 92 48 00
Germany
+49 6103 408 1014
India
+91 22 6742 9494
Italy
+39 02 950 591
Japan
+81 45 453 9100
Latin America
+1 561 688 8700
Middle East
+43 1 333 50 34 0
Netherlands
+31 76 579 55 55
New Zealand
+64 9 980 6700
Norway
+46 8 556 468 00
Russia/CIS
+43 1 333 50 34 0
Singapore
+65 6289 1190
Spain
+34 914 845 965
Sweden
+46 8 556 468 00
Switzerland
+41 61 716 77 00
UK
+44 1442 233555
USA
+1 800 532 4752
www.thermoscientific.com©2014 Thermo Fisher Scientific Inc. All rights reserved. ISO is a trademark of the International Standards Organization. All other trademarks
are the property of Thermo Fisher Scientific Inc. and its subsidiaries. This information is presented as an example of the capabilities of Thermo
Fisher Scientific Inc. products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property
rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your
local sales representative for details.
Thermo Fisher Scientific,
San Jose, CA USA is
ISO 13485 Certified.
Tacrolimus
Sirolimus
Cyclosporin A
Spiked
(ng/mL)
Measured
[average ± st dev]
(ng/mL)
Spiked
(ng/mL)
Measured
[average ± st dev]
(ng/mL)
Spiked
(ng/mL)
Measured
[average ± st dev]
(ng/mL)
Low Control
6.0
6.2 ± 0.2
3.6
3.2 ± 0.3
76
74 ± 2
Mid Control
12
13.1 ± 0.5
9.7
11.0 ± 0.4
199
175 ± 3
High Control
22
21.2 ± 2.8
17.4
17.6 ± 4.2
311
276 ± 9
Subject 1
10
11.0 ± 1.0
5.3
6.2 ± 0.5
0
< LOQ
Subject 2
4.2
4.9 ± 0.4
0
< LOQ
0
< LOQ
Subject 3
0
< LOQ
2.1
3.0 ± 0.9
59
57 ± 3