Background Image
Table of Contents Table of Contents
Previous Page  293 / 302 Next Page
Information
Show Menu
Previous Page 293 / 302 Next Page
Page Background

3

Results and Discussion

Method Productivity & Performance

The goal of the developed method was to decrease the

total work required by the laboratory for the analyses of

so many compounds. For this entire list of compounds to

be analyzed effectively, the TRACE 1310 GC and

TG-XLBMS column were used to optimize the

chromatographic separation of critical isomer pairs.

The chromatographic performance was such that all

compounds eluted within 17 minutes. The same separations

were observed after more than 100 injections of water,

soil, and building material extracts.

b

e

c

f

a

d

Same compounds 100

injections later

Figure 2: Chromatograms of several critical separations in a standard at 2000 pg absolute injection; except for benzo(b) and (k)

fluoranthene that are depicted in building material with a concentration of 400 pg

a: phenanthrene and anthracene

b: chrysene and benzo(a)anthracene

c: benzo(b) and benzo(k)fluroanthene

d: indeno (1,2,3,c,d)pyrene and dibenzo(a,h) anthracene

e: o,p DDD and p,p DDT

f: benzo(b) and benzo(k)fluoranthene in building material

after 100 injections of samples