Background Image
Table of Contents Table of Contents
Previous Page  289 / 302 Next Page
Information
Show Menu
Previous Page 289 / 302 Next Page
Page Background

RT:

6.57 -18.36

SM:

7B

7

8

9

10

11

12

13

14

15

16

17

18

Time (min)

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

RelativeAbundance

70ug/m3

Av.Ht.=1447020

28ug/m3

Av.Ht.=802015

14ug/m3

Av.Ht.=470651

5.6ug/m3

Av.Ht.=228736

2.8ug/m3

Av.Ht.=159554

Background

Av.Ht.=91223

NL:1.68E6

TIC F: -cAPCI

sid=5.00 SRMms2

126.000@cid1

3.00

[92.500-93.500] MS

Phenolcourb

PHENOL

RT

Syringe

Speed

Aver Height

Background

Subtracted

6.0

OFF

57639

4.6

249205

191566

Phenol

RT (min)

Syringe

Speed

Average Height

Background

Subtracted

Response/ISTD

Ratio

Concentration

(µg/m

3

)

7.4 - 8.2

9

1447020

1355797

7.1

70

9.1 - 9.9

7

802015

710792

3.7

28

10.9 - 11.7

5

470651

379428

2.0

14

12.6 - 13.4

3

228736

137513

0.7

5.6

14.4 - 15.2

1

159554

68331

0.4

2.8

16.0 - 16.8

OFF

91223

0

0.0

0

Std. Dev. = 3871

(3 x Std. Dev.) =

11613

0.06

0.4

LOD

ISTD: Acetic acid

13

D

2

, syringe 1 mL, Speed 2

SRM (

m/z

94

m/z

61)

SRM (

m/z

126

m/z

93)

Calibration Curve

PHENOL

y = 0.6023x

2

+ 5.5954x

R

2

= 0.999

0

10

20

30

40

50

60

70

80

0.0

1.0

2.0

3.0

4.0

5.0

6.0

7.0

8.0

Conc.ug/m

3

Ratio (

m/z

126

m/z

93)/ISTD

Figure 10: Calibration curve for phenol.

Figure 9: Reconstructed ion trace for phenol to produce the calibration curve.

Table 8. Quantitative results for phenol, and LOD determination.