Background Image
Previous Page  52 / 144 Next Page
Information
Show Menu
Previous Page 52 / 144 Next Page
Page Background

52

Pesticides in Water

Using 1g 6mL HyperSep C18 Extraction Column

(Part Number: 60108-301)

Sample Preparation

None

Condition C18 Extraction Column

10mL of hexane/acetone (50:50)

10mL of methanol

10mL of DI H

2

O

Apply Sample

Load 1 liter of sample at a rate of 8 to 10mL/minute

Wash Column

20mL of DI H

2

O

Dry Column

Use maximum vacuum pressure for 15 to 30 minutes

Elute Chlorophenoxy Acid Herbicides

10mL of hexane/acetone (50:50)

Concentration/Evaporation

Add 500µL of a keeper solvent (methanol, DMF, other)

Evaporate to 500µL under a nitrogen stream at room

temperature

Pesticides Extracted

α

-hexachlorocychexane

4,4'-DDE

Lindane

Dieldrin

β

-hexachlorocychexane

Endrin

Heptachlor

4,4'-DDD

δ

-hexachlorocychexane

Endosulfan II

Aldrin

4,4'-DDT

Heptachlor

Endrin aldehyde

Endosulfan 1

Enrin sulfate

Environmental

Applications

GC/MS Determination of Phenols in Drinking Water

Using 500mg 6mL HyperSep Retain PEP Extraction Column

(Part Number: 60108-732)

Sample Preparation

Collect 1L of H

2

O

Adjust pH to 2 with 6N HCl

Condition HyperSep Retain PEP Extraction Column

1 x 3mL CH

3

Cl

1 x 3mL CH

3

OH

1 x 3mL 0.05N HCl

NOTE:

Do not allow the cartridge to dry out

Apply Sample

Load 1L of water sample at 20mL/minute

Dry column for 10 to 15 minutes

Wash Column

1 x 10mL H

2

O

Dry column (1 minutes at >10"Hg)

Elute

1 x 10mL CH

3

Cl

1 x 3mL CH

3

Cl

Concentrate the extract to 0.9mL in water bath (40°C)

under a gentle stream of nitrogen

Analysis

Adjust final volume to 1.0mL with CH

3

Cl

Analyze the extract with GC/MS

Recommended GC Column

Part Number

TraceGOLD TG-5MS 30m x 0.25m x 0.25µm

26098-1420