Background Image
Previous Page  56 / 144 Next Page
Information
Show Menu
Previous Page 56 / 144 Next Page
Page Background

56

Polynuclear Aromatic Hydrocarbons in Pond Water

Using 500mg 3mL HyperSep NH

2

Extraction Column

(Part Number: 60108-518)

Sample Preparation

Filter water sample through a 0.45µm filter

Add 2mL of methanol to 200mL of filtered water sample

Mix and degas sample for 2 minutes

Condition NH

2

Extraction Column

15 to 20mL of methylene

chloride/trichlorotrifluoroethylene (TCTFE)

15 to 20mL of TCTFE

Dry for 5 minutes

15 to 20mL of methanol

20mL of DI H

2

O

Apply Sample

Load 200mL of sample at a rate of 8 to 10mL/minute

Wash Column

20mL of DI H

2

O

Dry Column

Use maximum vacuum pressure for 15 to 30 minutes

Elute Polynuclear Aromatic Hydrocarbons

20mL of TCTFE

Concentration/Evaporation

Evaporate to dryness under a nitrogen stream at room

temperature

Injection/Analysis

Reconstitute with 100µL of TCTFE and inject 1 to 2µL

aliquot onto GC

Polynuclear Aromatic Hydrocarbons Extracted

Naphthalene

Chrysene

Fluorene

B(e)pyrene

Acenaphthene

B(b)fluoranthene

Phenanthrene

B(k)fluoranthene

Anthracene

B(a)pyrene

Fluoranthene

D(a,h)anthracene

Pyrene

B(g,hi)perylene

B(a)anthracene

Indeno(1,2,3-cd)pyrene

Praziquantel in Water

Using 60mg 3mL HyperSep Retain PEP Extraction Column

(Part Number: 60107-203)

Sample Preparation

To 10mL of water/methanol (95:5, v/v) containing

10ppm atrazine, add 20µL of acetic acid

Condition Retain PEP Extraction Column

2mL of methanol

2mL of DI H

2

O

Apply Sample

Load 4mL samples at 1 to 2mL/minute

Wash Column

1mL of methanol/DI H

2

O (5:95, v/v)

Elute Praziquantel

4mL of methanol

Analysis

Mobile phase: ACN/water (40:60, v/v)

Flow: 1mL/minute

Temperature: 30°C

Detection: UV 214nm

Recommended HPLC Column

Part Number

Hypersil GOLD 5µm, 250 x 4.6mm

25005-254630

Environmental

Applications