Background Image
Previous Page  61 / 144 Next Page
Information
Show Menu
Previous Page 61 / 144 Next Page
Page Background

61

Forensic

Applications

Amphetamines, Opiates and Phencyclidine in Oral Fluid

for GC/MS Confirmations

Using 50mg 1mL HyperSep Verify-CX Extraction Column

(Part Number: 60108-741)

Prepare Sample

Add 100 to 500µL of neat sample to a clean test tube

Add internal standard(s) and let sit for 10 minutes at

room temperature

Add 800µL of 100mM phosphate buffer (pH 6.0)

Mix/vortex for 10 seconds

Sample pH should be 6.0±0.5

Adjust pH accordingly with 100mM monobasic or dibasic

sodium phosphate

Condition Verify-CX Extraction Column

200µL of CH

3

OH

200µL of DI H

2

O

200µL of 100mM phosphate buffer (pH 6.0)

Apply Sample

Load sample at 1mL/minute (do not exceed this

flow rate)

Wash Column

500µL of DI H

2

O

500µL of 100mM acetic acid

500µL of CH

3

OH

Dry column (5 minutes at >10"Hg)

Elute Compounds

800µL of CH

2

Cl

2

/IPA/NH

4

OH (70:26:4)

Collect eluate at 1mL/min (do not exceed this rate)

NOTE:

Prepare elution solvent fresh daily; add IPA/NH

4

OH mix,

then add CH

2

Cl

2

(pH 11-12)

Dry Eluate

For amphetamines and PCP, add 100µL of 5% trifluo-

roacetic acid in methanol after 5 minutes of drying

(5 minutes drying removes ammonia, addition of acid

ionizes volatile analytes preventing loss)

Evaporate to full dryness at <40°C under a stream

of nitrogen

Derivatize

For Amphetamines*:

Add 50µL PFPA (PFAA)** then vortex. Overlay with

nitrogen and cap. React for 20 minutes at 70°C.

Evaporate to dryness at <40°C. Reconstitute with

50µL ethyl acetate.

For Opiates*:

Add 200µL of a 1:1 solution of PFPA (PFAA)**

(10 x 1mL ampules, Cat. No. TS-65193)

Vortex then react for 60 minutes at 40°C

Reconstitute with 50µL ethyl acetate

Quantitate

Inject 2µL onto gas chromatograph

* Alternative derivatizations may be used

** Part number TS-65193 (10 x 1mL ampules)

NOTE:

Phencyclidine does not derivatize

Recommended GC Columns

Part Number

TRACE TR-DoA5 Column, 15m x 0.25mm x 0.25µm

26AF130P

TRACE TR-DoA35 Column, 15m x 0.20mm x 0.33µm

26AC497P