Nitrobenzene in Water

Alternative Derivatization

Using 500mg 6mL HyperSep Retain PEP Extraction Column

(Part Number: 60107-206)

Sample Preparation

Add CH

3

OH to water sample to provide a 0.5% solution

Adjust pH to 7

Condition HyperSep Retain PEP Extraction Column

1 x 3mL n-hexane

1 x 5mL CH

3

OH

NOTE:

Do not allow the cartridge to dry out

Apply Sample

Load 1 x 10mL H

2

O

Elute at 5mL/minute

Wash Column

1 x 10mL H

2

O

Dry column (20 minutes at >10"Hg)

Elute

1 x 10mL n-hexane/acetone (90:10, V/V)

Concentrate sample to 1mL

Analysis

Inject onto HPLC

Recommended HPLC Column

Part Number

Hypersil GOLD 3µm, 150 x 4.6mm

25003-154630

Organochlorine Pesticides and PCB Extraction

Using 10g 75mL HyperSep C18 Extraction Column

(Part Number: 60108-703)

Sample Preparation

Adjust pH of sample to 2 using sulfuric acid

Add 1 x 5mL CH

3

OH

Condition HyperSep C18 Extraction Column

1 x 10mL methylene Chloride

1 x 10mL acetone

1 x 10 CH

3

OH

1 x 20mL DI H

2

O

NOTE:

Do not allow the cartridge to dry out

Apply Sample

Load sample at 1 to 3mL/minute

Wash Column

1 x 5mL acetone

Dry column (1 minutes at >10"Hg)

Elute

3 x 10mL CH

3

Cl

Dry the extract by passing it through anhydrous

sodium sulfate

Rinse the collection device with CH

3

Cl

Add the solvent to the sodium sulfate

Analysis

Concentrate the extract

Analyze as appropriate via GC/MS

Recommended GC Column

Part Number

TraceGOLD TG-OCP I 30m x 0.25m x 0.25µm

26078-1420

Clean-Up of Organochlorine Pesticides and PCB Extracts

Using 1g 6mL HyperSep Florisil Column

(Part Number: 60108-431)

Sample Preparation

Add hexane to sample

Condition HyperSep Florisil Extraction Column

1 x 9mL 90:10 hexane/acetone

Apply Sample

Load 2mL of sample at 1 to 2mL/minute

Elute

1 x 9mL of 90:10 hexane/acetone

Collect eluate at 1 to 2mL/minute

Evaporate

Evaporate eluates to 1mL under a gentle stream of

nitrogen <40°C

Reconstitute

Reconstitute sample to a final volume of 2mL with hexane

Environmental

Applications