Background Image
Previous Page  47 / 144 Next Page
Information
Show Menu
Previous Page 47 / 144 Next Page
Page Background

47

Ethametsulfuron in Solution

Using 60mg 3mL HyperSep Retain-AX Extraction Column

(Part Number: 60107-403)

Sample Preparation

Prepare a 10ppm solution of nicosulfuron diluted in 2%

ammonia aqueous solution

Condition Retain-AX Extraction Column

1mL of methanol

1mL of DI H

2

O

Apply Sample

Load 2mL samples

Wash Column

1mL of 2% ammonia hydroxide

1mL of methanol

Elute Ethametsulfuron

2mL of methanol with 2% acetic acid

Analysis

Mobile phase: methanol:2% acetic acid (60:40, v/v)

Flow: 1mL/minute

Temperature: 30°C

Detection: UV 355nm

Recommended HPLC Column

Part Number

Hypersil GOLD aQ 5µm, 250 x 4.6mm

25305-254630

Glyphosate and Glufosinate by Solid-Phase Anion Exchange Extraction

Using 1g 6mL HyperSep Verify-AX Extraction Column

(Part Number: 60108-732)

Sample Preparation

1L H

2

O adjusted to pH 6

Condition HyperSep Verify-AX Extraction Column

1 x 5mL CH

3

OH

1 x 10mL DI H

2

O

Apply Sample

Load sample at 1 to 3mL/minute

Wash Column

1 x 10mL DI H

2

O

Dry column (10 minutes at >10"Hg)

Elute

1 x 4mL of 1 mol/L HCl/CH

3

OH (4/1)

Add eluant draw through at 1mL/minute

Evaporate

Evaporate under a gentle stream of nitrogen in a water

bath heated to 50°C

Analysis

Add 50µL of MTBSTFA* and 50µL of dimethylformamide

for derivatization

Sonicate for 2 minutes

Heat to 80°C for 30 minutes

Cool to room temperature

Inject onto GC/MS

* Part number TS-48920

Recommended GC Column

Part Number

TraceGOLD TG-17MS 30m x 0.25m x 0.25µm

26089-1420

Environmental

Applications