Background Image
Previous Page  45 / 144 Next Page
Information
Show Menu
Previous Page 45 / 144 Next Page
Page Background

45

EPA Method 535 – Analysis of Chloroacetanilide and

Acetamide Herbicide Degradates in Water

Using 500mg 6mL HyperSep Hypercarb Extraction Column

(Part Number: 60106-402)

Sample Preparation

250mL water collected

Condition HyperSep Hypercarb Extraction Column

1 x 20mL ammonium acetate/CH

3

OH

1 x 30mL H

2

O

1 x 3mL H

2

O to top of cartridge

Apply Sample

Load sample at 10 to 15mL/minute

Wash Column

1 x 5mL H

2

O

Dry column

Elute

Insert fresh collection tubes into manifold

1 x 15mL of ammonium acetate/CH

3

OH

Collect eluate at 5mL/minute

Evaporation

Evaporate eluates under a gentle stream of nitrogen in a

heated water bath 60 to 70°C

Analysis

To sample add 1mL 5mM ammonium acetate/CH

3

OH

Inject onto LC/MS/MS

Compound

MRM Transition

Propachlor OA

206/134

Flufenacet OA

224/152

Propachlor ESA

256/80

Flufenacet ESA

274/80

Dimethenamid OA

270/198

Dimethenamid ESA

320/80

Alachlor OA

264/160

Acetochlor OA

264/146

Alachlor ESA

314/80

Metolachlor OA

278/206

Acetochlor ESA

314/80

Metolachlor ESA

328/80

Dimethachlor ESA (sur)

300/80

Butachlor ESA (IS)

356/80

Recommended HPLC Column

Part Number

Hypersil GOLD 3µm, 150 x 2.1mm

25003-152130

Environmental

Applications