Background Image
Previous Page  42 / 144 Next Page
Information
Show Menu
Previous Page 42 / 144 Next Page
Page Background

42

Atrazine in Water

Using 60mg 3mL HyperSep Retain PEP Extraction Column

(Part Number: 60107-203)

Sample Preparation

To 10mL of water containing 10ppm atrazine, add

20µL of acetic acid

Condition Retain PEP Extraction Column

2mL of methanol

2mL of DI H

2

O

Apply Sample

Load 4mL samples at 1 to 2mL/minute

Wash Column

1mL of DI H

2

O

Elute Atrazine

5mL of methanol

Dry Eluate and Reconstitute

Evaporate to full dryness under a stream of nitrogen

Reconstitute the sample to 0.5mL with water

Analysis

Mobile phase: ACN/0.2% acetic acid (10:90, v/v)

Flow: 1mL/minute

Temperature: 30°C

Detection: UV 214nm

Recommended HPLC Column

Part Number

Hypersil GOLD 5µm, 250 x 4.6mm

25005-254630

Bentazone in Water

Using 500mg 6mL HyperSep Retain PEP Extraction Column

(Part Number: 60107-206)

Sample Preparation

Adjust the pH of 500mL of water to

3 using

0.5 M H

2

SO

4

Condition Retain PEP Extraction Column

3mL of tetrahydrofuran

5mL of methanol

5mL of DI H

2

O

Apply Sample

Load the 500mL sample at a rate no greater than

5mL/min

Wash Column

5mL of DI H

2

O

Dry column (20 minutes at >10"Hg/full flow for positive

pressure manifold)

Elute Bentazone

Use 0.8mL of methanol to replace the residual water in

the PEP packing

Discard the water which passes through the SPE column

Wait for 2 minutes to make sure the methanol infiltrates

the packing material

Elute the column with 3mL of tetrahydrofuran

Collect the eluate and reconstitute to 3mL using

mobile phase

Environmental

Applications

Bentazone in Water

Alternative Derivatization

Using 500mg 6mL HyperSep Retain PEP Extraction Column

(Part Number: 60107-206)

Sample Preparation

Adjust pH to 3 with H

2

SO

4

Condition HyperSep Retain PEP Extraction Column

1 x 5mL THF

1 x 5mL CH

3

OH

1 x 5mL H

2

O

NOTE:

Do not allow the cartridge to dry out

Apply Sample

Load 1 x 500mL H

2

O

Elute at 5mL/minute

Wash Column

1 x 5mL H

2

O

Dry column (20 minutes under N

2

)

Elute

1 x 3mL THF

Concentrate sample to 3mL under N

2

Analysis

Inject onto HPLC