Background Image
Previous Page  43 / 144 Next Page
Information
Show Menu
Previous Page 43 / 144 Next Page
Page Background

43

Chlorophenoxy Acid Herbicides in Water

Using 1g 6mL HyperSep C18 Extraction Column

(Part Number: 60108-301)

Sample Preparation

Adjust pH of 1L of water sample to pH 1.0 with

hydrochloric acid

Condition C18 Extraction Column

10mL of hexane/acetone (50:50)

10mL of acidified methanol (5% HCl in methanol)

10mL of DI H

2

O

Apply Sample

Load 1 liter of sample at a rate of 8 to 10mL/minute

Wash Column

10mL of DI H

2

O adjusted to pH 1.0 with HCl

Dry Column

Use maximum vacuum pressure for 15 to 30 minutes

Elute Chlorophenoxy Acid Herbicides

10mL of hexane/acetone (50:50)

Concentrate/Evaporate

Add 500µL of a keeper solvent (methanol, DMF, other)

Evaporate to 500µL under a nitrogen stream at room

temperature

Injection/Analysis

Reconstitute with 100µL of TCTEF and inject at

1 to 2µL onto GC column

Chlorophenoxy Acid Herbicides Extracted

2,4-D acid

2,4,5-trichloro phenoxy propionic acid (Silvex)

Dicamba

Dinitro-sec-butyl phenol

Environmental

Applications

Chlorophenol in Water

Using 500mg 6mL HyperSep Retain PEP Extraction Column

(Part Number: 60107-206)

Sample Preparation

Collect 500mL of H

2

O

Adjust pH to 2 with H

2

SO

4

Condition HyperSep Retain PEP Extraction Column

1 x 5mL CH

3

OH

1 x 5mL H

2

O

NOTE:

Do not allow the cartridge to dry out

Apply Sample

Load 1 x 500mL H

2

O

Elute at 5mL/minute

Wash Column

1 x 5mL H

2

O

1 x 1mL CH

3

OH

Elute

1 x 3mL THF

Concentrate sample to 3mL under N

2

Analysis

Inject onto HPLC

Recommended HPLC Column

Part Number

Hypersil GOLD PFP 3µm, 150 x 4.6mm

25403-154630

2,4-Dichlorophenol in Water

Using 500mg 6mL HyperSep Retain PEP Extraction Column

(Part Number: 60107-206)

Sample Preparation

Adjust pH to 2 with H

2

SO

4

Condition HyperSep Retain PEP Extraction Column

1 x 5mL CH

3

OH

1 x 5mL H

2

O

NOTE:

Do not allow the cartridge to dry out

Apply Sample

Load 1 x 500mL H

2

O

Elute at 5mL/minute

Wash Column

1 x 5mL H

2

O

1 x 1mL CH

3

OH

Dry column (20 minutes under N

2

)

Elute

1 x 3mL THF

Concentrate sample to 3mL under N

2

Analysis

Inject onto HPLC

Recommended HPLC Column

Part Number

Hypersil GOLD PFP 3µm, 150 x 4.6mm

25403-154630