Background Image
Table of Contents Table of Contents
Previous Page  456 / 658 Next Page
Information
Show Menu
Previous Page 456 / 658 Next Page
Page Background

2

Quantitative Confirmatory Analysis of the NIDA 5 Panel Using Prelude SPLC System and TSQ Quantum Ultra MS

Overview

Purpose

: Develop and validate a simple and efficient quantitative LC-MS/MS

method for SAMHSA-compliant confirmatory analysis of 5 panel drug using

novel HPLC system.

Methods

: Human urine containing the drugs were spiked with internal

standards, enzymatically hydrolyzed, and diluted.

Results

: The LC-MS/MS method was developed and validated to comply with

SAMHSA guidelines.

Introduction

Effective on October 2011, the new SAMHSA/NIDA guidelines allow

implementation of LC-MS technique to perform NIDA-5 panel, urine quantitative

confirmatory analysis. LC-MS/MS methods are often less complicated than the

previously implemented GC-MS/MS methods because they do not require

derivatization. The NIDA-5 panel requires 6 separate quantitative methods for

analysis of THCA, opiates, amphetamines, cocaine, phencyclidine and 6-MAM to

confirm immunomethod positive samples. Here we developed 6 methods using

a single sample preparation procedure, analytical column, mobile phase and

instrument configuration. The methods are implemented on new Thermo

Scientific™ dual channel Prelude ™ SPLC online sample preparation-liquid

chromatography system, which allows method execution in parallel with a

different method on each channel or the same method on both channels

multiplexed to a single mass spectrometer.

Serial MS detection of multiplexed methods improves mass spectrometer

utilization time, increases laboratory throughput and reduces analysis cost. The

syringe pumps and high-pressure, low-volume gradient mixing used in the

Prelude SPLC system provide enhanced LC performance including improved peak

shape and resolution, stable retention times and reduced solvent consumption.

Methods

Sample Preparation

The sample prep procedure includes glucuronide hydrolysis followed by dilution.

For each sample a 200-µL aliquot of urine was spiked with 10 µL of internal

standard solution and 100 µL of β-glucuronidase enzyme in ammonium acetate

buffer, pH=5.0. The samples were incubated at 60 °C for 2 hours. A 200-µL

aliquot of methanol was added to each sample to stop enzymatic reaction.

Samples were cooled down, centrifuged and diluted 20-fold with water, except

for THCA, which was diluted 2-fold with water. Then 20 µL of sample was

injected onto the LC-MS/MS system.

Liquid Chromatography

Chromatographic separations were performed with the Prelude SPLC system by

direct injection onto a Thermo Scientific™ Accucore™ PFP 50x2.1mm, 2.6 µm

analytical column. The column was maintained at room temperature. Mobile

phases A and B consisted of 10 mM ammonium formate with 0.1% formic acid in

water and methanol, respectively. Separate methods were set up to analyze 6-

MAM, BE, PCP, and THCA. One method was set up for the combination of

amphetamine, methamphetamine, MDA, MDEA and MDMA. A final method was

used for the opiates morphine and codeine along with hydromorphone,

hydrocodone, oxymorphone and oxycodone. Figure 1 shows the LC method for

analyzing the opiates.

Mass Spectrometry

MS/MS analysis was carried out on a Thermo Scientific™ Quantum Ultra™ triple

quadrupole mass spectrometer equipped with a heated electrospray ionization

(HESI-II) probe. MRM transitions for each compound are listed in Table 1.

Validation

The calibration standards and quality control (QC) samples were prepared by

spiking compounds into blank urine. Samples were processed as described in

the Sample Preparation section. Methods were validated in multiplexed mode.

Intra- and inter- method precision and accuracy were determined by analyzing a

calibration curve along with replicate QCs on three different days. Matrix effects

were determined by comparing peak area of samples processed in multiple lots

of urine to that of one process in water. Additionally for the opiates, we were

able to correlate results obtained with this method to those from a toxicology

laboratory validated method.

Data Analysis

Thermo Scientific™ TraceFinder™ software was used for data acquisition and

processing. Data were processed with ion ratio confirmation.

Results

For each method, performance wa

quantitation limits (LOQ) for some

demonstrate method capability. T

PCP and THCA; 5-2000 ng/mL for

10-2000 ng/mL for morphine, cod

(Figure 2). The intra-method preci

9.6%, <15.9% for PCP, BE, 6-MAM,

respectively. The inter-method pre

<7.0%, <15.3% for PCP, BE, 6-MA

respectively. These results are su

were seen and those were largely

The percent recovery for 8 spiked

80-120% (Table 3). Data collected

correlated well with toxicology lab

>0.99 (Figure 4). Implementation

with syringe pumps improved rete

shape and resolution, thus allowin

methods while still keeping good

FIGURE 1. LC method for separat

TABLE 1. List of NIDA 5 compou

LOQ and Linear range

Drug

MRM (Q: Quan

Amphetamine 136.1-91.3 (Q), 13

Methamphetamine 150.2-91.2 (Q), 15

MDA

180.2-135.2 (Q), 1

MDMA

194.1-163.1 (Q), 1

MDEA

208.1-163.1 (Q), 2

Benzoylecgonine 290.1-168.1 (Q), 2

THCA

354.3-336.3 (Q), 3

Phencyclidine 244.2 -159.1 (Q), 2

Morphine

286.11-152.

286.11-16

Codeine

300.2-152.1 (Q), 3

6-Acetylmorphine 328.1-165.1 (Q), 3