Background Image
Table of Contents Table of Contents
Previous Page  418 / 658 Next Page
Information
Show Menu
Previous Page 418 / 658 Next Page
Page Background

2

Verification of an LC-MS/MS Forensic Method for 19 Opioids, Opiates, and Their Metabolites in Human Urine Without Hydrolysis

Overview

Purpose:

This is the verification of a forensic method containing a panel of 19 opiates

without the labor-intensive step of hydrolysis. These compounds were analyzed using

a Thermo Scientific™ Prelude SPLC™ system for chromatographic separation coupled

with a Thermo Scientific™ TSQ Endura™ triple quadrupole mass spectrometer.

Methods:

All compounds were spiked into human urine, diluted, and injected onto the

Prelude SPLC system equipped with a Thermo Scientific™ Accucore™

aQ analytical

column. After elution from the analytical column, the compounds were simultaneously

analyzed using the TSQ Endura MS with a heated electrospray ionization (HESI-II)

probe in positive mode. The total method time was approximately four minutes,

allowing for rapid quantification of all analytes and 10 internal standards.

Results:

All 19 opiates, opioids, and their metabolites were verified successfully and

passed all acceptance criteria. These compounds were analyzed without hydrolysis,

drastically reducing the total sample preparation time.

Introduction

It is currently common practice in many forensic laboratories to analyze opiates and

opioids using hydrolysis during sample preparation. This process can take up to

24 hours to complete so that all conjugated metabolites are converted back to their

parent molecules. We introduce a method that alleviates these time constraints,

allowing for an efficient, sensitive analysis of the intact metabolites. Such metabolites

(for example, morphine glucuronides) have a reputation for being difficult to

chromatograph and detect at low levels. The Prelude SPLC system equipped with an

Accucore aQ column offers great retention of all analytes and resolution of isobars.

The highly sensitive TSQ Endura MS reproducibly detects the low end of the

calibration range for even the least responsive analytes and provides accurate

quantification of these somewhat problematic compounds.

Methods

Sample Preparation

The analytes of interest were spiked into human urine at various concentrations to

make calibrators and controls. Each sample was divided into two parts; one set was

hydrolyzed prior to analysis and the second set was analyzed directly. Hydrolysis was

performed by adding 1 M ammonium acetate buffer containing

b

-glucuronidase to

samples and incubating overnight at 60

°

C. The nonhydrolyzed samples had the same

volume of ammonium acetate buffer added but without any

b

-glucuronidase present

and no incubation. Isotopically labeled internal standards were then added to all the

samples. The samples were vortexed and centrifuged. The supernatant was removed

from the pellet and transferred into clean vials for LC-MS/MS analysis.

Liquid Chromatography

The LC method is shown in Table 1. A Prelude SPLC system, seen in Figure 1, was

used in LX mode, equipped with a 2.1 x 100 mm, 2.6 µm particle size Accucore aQ

analytical column. The mobile phases consisted of 0.1% formic acid in (A) water and

(B) methanol. Using less than 2 mL of solvent, per injection, the LC system was able to

successfully resolve all isobaric compounds within this method. The total method time

of 4.25 min, when multiplexed, allows for results every 2 min for all 19 compounds.

Step

Start

(min)

Time

(s)

Flow

(mL/min)

Grad

%A

%B

1

0.00

20

0.40

Step

100.0

-

2

0.33

5

0.40

Step

92.0

8.0

3

0.42

50

0.40

Step

92.0

8.0

4

1.25

5

0.40

Step

75.0

25.0

5

1.33

130

0.40

Ramp

65.0

35.0

6

3.50

45

0.40

Step

-

100.0

7

4.25

100

0.40

Step

100.0

-

TABLE 1. Liquid chromatography program for the method, including solvent

composition, flow rate, and timing

FIGURE 1. Prelude S

Mass Spectrometry

A TSQ Endura triple q

positive mode was use

Data Analysis

All data acquisition an

Scientific™ Aria™ MX

software version 3.1.

Results

Calibration standards co

5

500 ng/mL were prep

prepared in urine at thre

tested by using five repli

quantitating them using

Carryover was calculate

quantitation (LLOQ) by t

limit of quantitation (UL

signal. Additionally, auto

QC samples that were r

to a freshly prepared cal

The assay precision ha

calibration standard lev

calibrations standard lev

passed acceptance crite

chromatograms at the lo

in Figure 2. Additionally,

glucuronides in Figure 3

for oxymorphone-3

b

-glu

R

2

> 0.9900.

Lastly, all mass spectro

the Thermo Scientific™

transitions and paramet

spectrometers required

analysis.