6
Improving Throughput for Highly Multiplexed Targeted Quanti cation Methods Using Novel API-Remote Instrument Control and State-Model
Data Acquisition Schemes
Conclusion
The developments here resulted in the successful qualitative/quantitative analysis for
over 3,000 peptides representing over 2,000 proteins in this complex leukemia cell
digest. Successful quantification was determined for proteins spiked at over a 20-fold
range and the ability to change instrument acquisition parameters for increased
sensitivity.
nge
ide list
in each retention time window
l line and imported into the new
rary information to select unique
nformation used to perform real-time
00 peptides were chosen and 20-fold
:10 could not be calculated using the
in tandem MS/MS scan (panel B, and
FIGURE 6. The benefit of MS/MS scan (with higher S/N) compared to full scan.
Ratio of 1:10 could not be calculated in full scan (panel A), but it could be
calculated in tandem MS/MS scan (panel C).
All trademarks are the property of Thermo Fisher Scientific and its subsidiaries.
This information is not intended to encourage use of these products in any manners that might infringe the
intellectual property rights of others.
Light Channel
Heavy Channel
A
Zoom in
Product ion from heavy
Product ion from light
B
C