Application Note 613
Conclusion
In this study, the utility of direct dilute-and-shoot analysis
of glucuronic acid conjugates of stanozolol by means
of liquid chromatography and high-resolution,
accurate-mass mass spectrometry in sports antidoping
testing was assessed and demonstrated. Characterized and
validated by means of commercially available
3’-OH-stanozolol glucuronide, dilute-and-inject and
confirmation methods were established, which allowed for
the unambiguous identification of stanozolol misuse in
routine doping control samples. Additionally, new
long-term metabolites for the detection of stanozolol
abuse were observed in administration study urine
samples. These new target analytes, assigned as
stanozolol-
N
-glucuronide and 17-epistanozolol-
N
-
glucuronide, were characterized by mass spectrometry,
and hydrolysis experiments. Both proved particularly
useful as target compounds, enabling the determination of
drug abuse for up to 28 days post-administration of 5 mg
of stanozolol. Since high-resolution, accurate-mass mass
spectrometry has been found to be essential for the
successful identification of lowest amounts of stanozolol
metabolites in human urine, expansion of its use in
doping control is encouraged.
Reference
1. Schänzer, W.; Guddat, S.; Thomas, A.; Opfermann, G.;
Geyer, H.; Thevis, M. Expanding analytical possibilities
concerning the detection of stanozolol misuse by means
of high resolution/high accuracy mass spectrometric
detection of stanozolol glucuronides in human sports
drug testing,
Drug Test Anal.
2013
Nov-Dec; 5(11-12),
810-818.
2. Mareck, U.; Geyer, H.; Opfermann, G.; Thevis, M.;
Schänzer, W. Factors Influencing the Steroid Profile in
Doping Control Analysis,
J. Mass Spectrom.
2008
,
43
,
877-891.
AN64248-EN 1014S
Africa
+43 1 333 50 34 0
Australia
+61 3 9757 4300
Austria
+43 810 282 206
Belgium
+32 53 73 42 41
Canada
+1 800 530 8447
China
800 810 5118
(free call domestic)
400 650 5118
Denmark
+45 70 23 62 60
Europe-Other
+43 1 333 50 34 0
Finland
+358 9 3291 0200
France
+33 1 60 92 48 00
Germany
+49 6103 408 1014
India
+91 22 6742 9494
Italy
+39 02 950 591
Japan
+81 45 453 9100
Korea
+82 2 3420 8600
Latin America
+1 561 688 8700
Middle East
+43 1 333 50 34 0
Netherlands
+31 76 579 55 55
New Zealand
+64 9 980 6700
Norway
+46 8 556 468 00
Russia/CIS
+43 1 333 50 34 0
Singapore
+65 6289 1190
Spain
+34 914 845 965
Sweden
+46 8 556 468 00
Switzerland
+41 61 716 77 00
UK
+44 1442 233555
USA
+1 800 532 4752
www.thermoscientific.com©2014 Thermo Fisher Scientific Inc. All rights reserved. Winstrol is a registered trademark of Sanofi-Synthelabo Inc.
Nucleodur is a registered trademark of trademark of Macherey, Nagel GMBH & Co. ISO is a registered trademark of the
International Organization for Standardization (Organisation Internationale De Normalization). All other trademarks are the
property of Thermo Fisher Scientific and its subsidiaries. This information is presented as an example of the capabilities of
Thermo Fisher Scientific products. It is not intended to encourage use of these products in any manners that might infringe
the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are
available in all countries. Please consult your local sales representative for details.
Figure 6. Chromatograms of an authentic doping control routine sample representing an adverse analytical
finding for stanozolol in both screening and confirmation assays.
Thermo Fisher Scientific, (Bremen) GmbH
Management System Registered is
ISO 9001:2008 Certified.