LC/MS/MS Research Method for 14 Antidepressants Utilizing Dried Blood Spots

Overview

Purpose:

To develop an LC-MS/MS research method for measuring the

concentration of antidepressants from dried blood spots (DBS).

Methods:

An LC-MS/MS method was developed to extract 14 antidepressants

from dried blood spots for quantitation.

Results:

The analytes of interest were extracted successful from dried blood

spots showing good linearity, accuracy, and precision.

Introduction

Important factors in the analysis of drugs in whole blood are accurate

measurements, storage capabilities, small sample volume, and easy extraction.

Dried blood spots are becoming an adopted clinical research technique for the

analysis of drugs in biological matrices. Due to the complexity of the solution

resulting from dissolving the blood spot, the sample must undergo further

cleanup by chromatographic separation before introduction into the mass

spectrometer. A research application is demonstrated using the Thermo

Scientific™ Prelude SPLC™ system and the Thermo Scientific™

TSQ Endura

™ triple

quadrupole mass spectrometer (Figure 1) to quantitatively

analyze 14 antidepressant drugs collected from dried blood spots.

Instrumentation

A Prelude SPLC system was used

Scientific™

TurboFlow

Fluoro X

2.1 x 50 mm, 2.6 µm particle size

column. The detector for the syst

spectrometer with a heated electr

mode. Thermo Scientific™

TraceF

quantitation. The liquid chromatog

Method Parameters

Mobile phases were (A) 10 mM a

(B) 10 mM ammonium formate, 0.

45/45/10 acetonitrile/isopropanol/a

The mass spectrometer quantifier

(SRM) transitions are shown in Ta

was 10

750 ng/mL.

FIGURE 1. Prelude SPLC system with

TSQ Endura MS

Methods

Sample Preparation

The analytes of interest were spiked into human whole blood at various

concentrations to make calibrators and controls. The samples were spotted at a

fixed volume onto Whatman

®

paper. Then, the analyte was extracted by solvent

containing isotopically labeled internal standards and transferred into clean vials

for LC-MS/MS analysis. The entire workflow is depicted in Figure 2.

SpotDBS card with

standardsandQCs

Extract solutio

wellsandadd

Punchcardwith

3mm holepunch

FIGURE 2. Workflow of preparin

FIGURE 3. Online sample cleanu