Sample #

Compounds identified

Compounds identified

using ToxSpec Analyzer

using REMEDi HS

1

Acetaminophen

Not detected

Nortriptyline

Not detected

Amitriptyline

Amitriptyline

Oxazepam

Not detected

2

Nordiazepam

Nordiazepam

Alprazolam

Not detected

Cyamemazine

Cyamemazine

3

Acetaminophen

Not detected

Nordiazepam

Nordiazepam

Venlafaxine

Venlafaxine

Oxazepam

Oxazepam

Alprazolam

Not detected

4

Nordiazepam

Not detected

Diazepam

Diazepam

Oxazepam

Not detected

Temazepam

Not detected

Levomepromazine

Levomepromazine

Zopiclone

Zopiclone

5

Oxazepam

Not detected

Clomipramine

Clomipramine

Quinidine

Quinine

Haloperidol

Not detected

Clonazepam

Not detected

6

Acetaminophen

Not detected

Bisoprolol

Bisoprolol

7

Venlafaxine

Venlafaxine

Risperidone

Not detected

8

Quinine

Quinine

Hydromorphone

Hydromorphone

Morphine

Morphine

9

Lidocaine

Not detected

Nortriptyline

Not detected

Mirtazapine

Not detected

Amitriptyline

Amitriptyline

Cyamemazine

Cyamemazine

Levomepromazine

Levomepromazine

Zopiclone

Not detected

10

Bromazepam

Bromazepam

Paroxetine

Not detected

11

Sertraline

Not detected

Hydrocortisone

Not detected

12

Acetaminophen

Not detected

Alprazolam

Alprazolam

Prednisolone

Not detected

Hydroxyzine

Hydroxyzine

Fexofenadine

Not detected

TOTAL

46 Molecules

22 molecules

Table 1: List of psychoactive molecules identified in real laboratory samples

using the ToxSpec Analyzer compared to the REMEDi HS system

Our aim was to quickly and confidently identify toxic

compounds in the samples by spectral library searching

while performing a semi-quantification calculation for

identified compounds. To perform the semi-quantification,

a response factor that correlated the intensity of the MS

2

spectra to a concentration was calculated for each

molecule present in the library using internal standards.

The semi-quantification result was automatically

calculated using ToxID software. An example of the

automatically-generated report can be seen in Figure 2.

The report includes a list of compounds identified in a real

laboratory sample and their respective calculated

concentrations.

One important aspect of this method is the ability to

reprocess data retrospectively from the MS spectra. The

ToxID report is based on MS

2

spectra library searching.

This means that if the entry corresponding to the

compound is not currently available in the library, ToxID

will not be able to identify the analyte. However, as data

are acquired in MS mode, it is possible to reprocess the

MS trace and check that all major ions have been

identified by ToxID. If not, it is then possible to re-inject

the sample and perform MS

2

acquisition on specific ions.

Conclusion

The ToxSpec Analyzer is a good replacement for the

REMEDi HS system in research toxicology laboratories

because it offers increased sensitivity, greater specificity,

and lower cost-per sample analysis.