Conclusion
The use of a simple rapid acetonitrile work-up followed by
a TurboFlow method (online extraction and chromatography)
on the Thermo Scientific Transcend TLX-1 system with
tandem MS/MS allowed the specific and sensitive analysis of
various common opiates and their metabolites from a small
volume of whole blood. Moreover, a limited portion of
the acetonitrile extract volume was utilized in the analysis,
thus, the method presents potential to scale down to a
volume of blood achievable from a finger prick (5–10 µL).
The calibration curves for all analytes analyzed were
linear over the concentration range and carryover was
calculated at less than 1% for all analytes. Since the
method is ~ 4 minutes, 15 samples per hour may be
completed, or indeed, doubled/quadrupled with the
use of multiplexing. Significant time is saved in the
absence of SPE sample preparation.
The method enables the forensic toxicologist to produce
a full picture of the opiates and metabolites in blood to assist
with the determination of time of injection (presence of
6-MAM) and the detection of M3G and M6G to determine
prior use or accumulation following heavy use.
References and Acknowledgements
1. Morgan O, Griffiths C, Toson B, Rooney C, Majeed A, Hickman M.
Trends in deaths related to drug misuse in England and Wales, 1993-2004.
Health Statistics Quarterly 2006; 31: 23-7.
(http://www.statistics.gov.uk/downloads/theme_health/HSQ31.pdf)2. Therapeutic Drugs, Second Edition, Ed C Dollery, Churchill & Livingstone,
London, 1998
3. Disposition of Toxic Drugs and Chemicals in Man. 8th Edition,
Ed Randall C Baselt. Biomedical Publications, Foster City, Ca, 2008
The authors would like to acknowledge LGC Forensics for the prepared
samples and Francois Espourteille for the original proof of concept work
used in the methodology described here.
In addition to these
offices, Thermo Fisher
Scientific maintains
a network of represen-
tative organizations
throughout the world.
Africa-Other
+27 11 570 1840
Australia
+61 2 8844 9500
Austria
+43 1 333 50 34 0
Belgium
+32 2 482 30 30
Canada
+1 800 530 8447
China
+86 10 8419 3588
Denmark
+45 70 23 62 60
Europe-Other
+43 1 333 50 34 0
Finland/Norway/
Sweden
+46 8 556 468 00
France
+33 1 60 92 48 00
Germany
+49 6103 408 1014
India
+91 22 6742 9434
Italy
+39 02 950 591
Japan
+81 45 453 9100
Latin America
+1 608 276 5659
Middle East
+43 1 333 50 34 0
Netherlands
+31 76 579 55 55
South Africa
+27 11 570 1840
Spain
+34 914 845 965
Switzerland
+41 61 716 77 00
UK
+44 1442 233555
USA
+1 800 532 4752
www.thermo.comAN63025_E 04/09M
Part of Thermo Fisher Scientific
Thermo Fisher Scientific,
San Jose, CA USA is ISO Certified.
Legal Notices
©2009 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries.
This information is presented as an example of the capabilities of Thermo Fisher Scientific Inc. products. It is not intended to encourage use of these
products in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change.
Not all products are available in all countries. Please consult your local sales representative for details.
View additional Thermo Scientific LC/MS application notes at:
www.thermo.com/appnotesFor research use only. Not for use in diagnostic procedures.