Background Image
Table of Contents Table of Contents
Previous Page  293 / 658 Next Page
Information
Show Menu
Previous Page 293 / 658 Next Page
Page Background

Conclusion

A robust method with simple and easy sample preparation

was developed and validated for forensic toxicology

laboratories. The data window and total run time make

Table 1. Inter-assay quality control statistics for validation runs

Methadone EDDP

Meperidine Normeperidine Propoxyphene Norpropoxyphene

LQC

4.16/4.71

2.67/6.04

0.493/5.93

7.04/9.17

5.45/3.42

3.28/8.64

MQC

7.32/2.47

-5.72/4.48

3.55/5.31

-0.747/7.61

4.36/5.28

0.933/3.88

HQC

10.9/2.69 -0.587/2.28

5.67/4.13

3.93/5.92

1.81/5.96

-3.77/8.18

%Bias/%CV

Table 2. Percent recovery of six synthetic opioids in eight lots of human urine

Compound

Lot A

Lot B

Lot C

Lot D

Lot E

Lot F

Lot G

Lot H

Methadone

105

92.5

97.3

103

98.5

98.5

99.3

98.4

EDDP

94.9

97.1

95.1

96.0

105

92.4

94.9

98.2

Meperidine

104

97.6

105

98.4

110

104

98.7

102

Normeperidine

117

115

111

94.7

105

111

108

118

Propoxyphene

98.7

97.3

96.4

99.8

98.7

89.1

95.5

102

Norpropoxyphene

97.8

92.2

87.3

101

94.5

96.4

97.1

104

% Recovery

Figure 2. SRM chromatogram of six synthetic opioids and metabolites in urine at 20 ng/mL

0.2

0.4

0.6

0.8

1.0

1.2

1.4

1.6

1.8

2.0

2.2

2.4

2.6

2.8

3.0

3.2

3.4

3.6

3.8

4.0

4.2

4.4

Time (min)

Meperidine

Normeperidine

Norpropoxyphene

Propoxyphene

EDDP

Methadone

Figure 2: Reconstructed SRM chromatogram of six synthetic opioids and metabolites in urine

this method amenable to multiplexing with the Thermo

Scientific Transcend LX-2 LC system. Multiplexing with

the Transcend™ LX-2 LC system would result in a run

time of 2.5 minutes per sample.