Part of Thermo Fisher Scientific
www.thermoscientific.comLegal Notices: ©2011 Thermo Fisher Scientific Inc. All trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. This information is
presented as an example of the capabilities of Thermo Fisher Scientific Inc. products. It is not intended to encourage use of these products in any manners that
might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please
consult your local sales representative for details.
Thermo Fisher Scientific,
San Jose, CA USA is ISO Certified.
AN63452_E 08/11S
In addition to these
offices, Thermo Fisher
Scientific maintains
a network of represen
tative organizations
throughout the world.
Africa-Other
+27 11 570 1840
Australia
+61 3 9757 4300
Austria
+43 1 333 50 34 0
Belgium
+32 53 73 42 41
Canada
+1 800 530 8447
China
+86 10 8419 3588
Denmark
+45 70 23 62 60
Europe-Other
+43 1 333 50 34 0
Finland/Norway/
Sweden
+46 8 556 468 00
France
+33 1 60 92 48 00
Germany
+49 6103 408 1014
India
+91 22 6742 9434
Italy
+39 02 950 591
Japan
+81 45 453 9100
Latin America
+1 561 688 8700
Middle East
+43 1 333 50 34 0
Netherlands
+31 76 579 55 55
New Zealand
+64 9 980 6700
Russia/CIS
+43 1 333 50 34 0
South Africa
+27 11 570 1840
Spain
+34 914 845 965
Switzerland
+41 61 716 77 00
UK
+44 1442 233555
USA
+1 800 532 4752
Conclusion
TraceFinder 1.1 software was effectively used to perform
routine analysis of the synthetic opiates in urine. The
software enabled easy method setup, batch creation and
submission, and real time monitoring. The data review
functionality was useful for quick review and verification
of the data. The generated reports had all the necessary
information for record keeping for forensic laboratories.
Results and Discussion
The method was linear from 20 to 5000 ng/mL for five of
the six compounds. Normeperidine was linear from 2 to
1000 ng/mL. Standard accuracy ranged between 87.3%
and 115%. Matrix effects were investigated by analyzing
QCs prepared from six different lots of blank human
urine. All samples showed recoveries within 20% at
50 ng/mL. The assay performance is summarized in
Table 1. Figure 10 shows the SRM chromatograms of all
six synthetic opiates at the limit of quantitation (LOQ).
Table 1. Assay performance for six synthetic opiates in urine
% Recovery
Lot 1
Lot 3
Lot 4
Lot 5
Lot 6
Lot 7
Lot 8
Lot 9
R
2
Linear Range
LOQ
Methadone
105.0% 92.5% 97.3% 103.0% 98.5% 98.5% 99.3% 98.4% 0.9945 20-5000 ng/mL 20
EDDP
94.9% 97.1% 95.1% 96.0% 105.0% 92.4% 94.9% 98.2% 0.9951 20-5000 ng/mL 20
Meperidine
104.0% 97.6% 105.0% 98.4% 110.0% 104.0% 98.7% 102.0% 0.9935 20-5000 ng/mL 20
Normeperidine
117.0% 115.0% 111.0% 94.7% 105.0% 111.0% 108.0% 118.0% 0.9998
2-1000 ng/mL
2
Propoxyphene
98.7% 97.3% 96.4% 99.8% 98.7% 89.1% 95.5% 102.0% 0.9994 20-5000 ng/mL 20
Norpropoxyphene-
dehydrate
97.8% 92.2% 87.3% 101.0% 94.5% 96.4% 97.1% 104.0% 0.9989 20-5000 ng/mL 20