Background Image
Table of Contents Table of Contents
Previous Page  278 / 658 Next Page
Information
Show Menu
Previous Page 278 / 658 Next Page
Page Background

Part of Thermo Fisher Scientific

www.thermoscientific.com

Legal Notices: ©2011 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries.

This information is presented as an example of the capabilities of Thermo Fisher Scientific Inc. products. It is not intended to encourage use of these products in

any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available

in all countries. Please consult your local sales representative for details.

Thermo Fisher Scientific,

San Jose, CA USA is ISO Certified.

In addition to these

offices, Thermo Fisher

Scientific maintains

a network of represen­

tative organizations

throughout the world.

Africa-Other

+27 11 570 1840

Australia

+61 3 9757 4300

Austria

+43 1 333 50 34 0

Belgium

+32 53 73 42 41

Canada

+1 800 530 8447

China

+86 10 8419 3588

Denmark

+45 70 23 62 60

Europe-Other

+43 1 333 50 34 0

Finland/Norway/

Sweden

+46 8 556 468 00

France

+33 1 60 92 48 00

Germany

+49 6103 408 1014

India

+91 22 6742 9434

Italy

+39 02 950 591

Japan

+81 45 453 9100

Latin America

+1 561 688 8700

Middle East

+43 1 333 50 34 0

Netherlands

+31 76 579 55 55

New Zealand

+64 9 980 6700

Russia/CIS

+43 1 333 50 34 0

South Africa

+27 11 570 1840

Spain

+34 914 845 965

Switzerland

+41 61 716 77 00

UK

+44 1442 233555

USA

+1 800 532 4752

AN63497 _E 11/11S

Table 2. Statistical results for HRAM LC/MS quantitation of diazepam in urine

Table 2 [Diaz pam, SIM]

Specified Amount

(ng /mL )

Mean Calculated Amt.

(ng/mL)

%Accuracy

%CV

0.0125

0.0113

90.1

6.0

0.0250

0.0236

94.3

7.9

0.0625

0.0610

97.6

1.4

0.1250

0.127

101.9

2.9

0.250

0.273

109.1

1.7

2.50

2.65

105.9

1.0

25.0

25.5

102.0

1.7

125.0

123.5

98.8

0.8

250.0

250.8

100.3

1.2

spectrometer using multiplexed SIM are comparable

to those observed on triple stage quadrupole mass

spectrometers in SRM mode.

an R

2

regression value of 0.9998 using 1/x weighting.

Table 2 presents the statistical results for the HRAM

quantification of diazepam. The quantitative accuracy

and precision values obtained by the Q Exactive mass

Conclusion

The Q Exactive HRAM LC/MS system is a powerful

and flexible instrument that can provide both sample

identification and quantitative information for forensic

toxicology with a single sample analysis. By using the

method of multiplexed SIM, eight benzodiazepines in

urine were quantified with LLOQs at the pg/mL level

and with linear dynamic ranges of 3 to 4 orders of

magnitude.

For Forensic Toxicology Use Only