Background Image
Table of Contents Table of Contents
Previous Page  28 / 302 Next Page
Information
Show Menu
Previous Page 28 / 302 Next Page
Page Background

flow rate of 850 µL/min, the 6-port valve was switched

back to the starting position. The loading and analytical

columns were cleaned with a high organic phase before

being re-equilibrated to their starting conditions. The total

run time for each analysis was six minutes. The mobile

phases for the analysis were water and acetonitrile, both

containing 0.1% formic acid. The gradient profile for each

pump is shown in Figure 2.

The pressure at the beginning of the gradient was

monitored. At a flow rate of 850 µL/min (at the initial

gradient conditions with the flow going through only the

Hypersil GOLD 50

×

2.1 mm, 1.9 µm column), the back-

pressure for the Fast-HPLC system was approximately

450 bar. For comparison, an earlier method which used

a Hypersil GOLD 50

×

2.1 mm, 3 µm column had a

backpressure of approximately 150 bar at a flow rate

of 200 µL/min.

MS

MS analysis was carried out on a TSQ Quantum Access

triple stage quadrupole mass spectrometer with a heated

electrospray ionization (H-ESI) probe (Thermo Scientific,

San Jose, CA). The MS conditions were as follows:

Ion source polarity: Positive ion mode

Spray voltage: 4000 V

Vaporizer temperature: 450°C

Sheath gas pressure (N

2

): 50 units

Auxiliary gas pressure (N

2

): 50 units

Ion transfer tube temperature: 380°C

Collision Gas (Ar): 1.0 mTorr

Q1/Q3 Peak Resolution: 0.7 u

Scan Width: 0.002 u

Results and Discussion

Chromatograms for the calibration standard at a con-

centration of 500 pg/mL are shown in Figure 3. In the

Fast-HPLC run, all 11 of the individual analytes were

eluted before three minutes. In contrast, none of the

analytes in the standard HPLC run were eluted until

nearly eight minutes into the run. Further optimization

of the chromatography for the Fast-HPLC would produce

even shorter run times.

Calibration curves for all 11 compounds were

generated using LC

QUAN

2.5 software (Thermo

Scientific, San Jose, CA). Excellent linearity was achieved

for all of the compounds analyzed in this experiment.

Figure 4 shows a representative calibration curve for the

compound azoxystrobin over the concentration range

0.5 to 1000 pg/mL (ppt). The calibration curve fit

parameters and the limits of detection for the analytes

are summarized in Table 2. The final column in the table

lists the Minimum Performance Reporting Limit (MPRL)

for these compounds as set by the Japanese Ministry of

Health, Labour, and Welfare

1

. All of the compounds

were detected and quantified at levels well below these

regulatory requirements.

Figure 2: Gradient profiles for the two LC pumps used in this experiment.

The Fast-HPLC pump gradient is shown on the left, and the loading pump

gradient is show on the right.

Figure 3: Chromatograms showing the SRMs for each of the components in

the mixture. Two different HPLC conditions are shown: the Fast-HPLC run

and the standard HPLC run. All compounds in the Fast-HPLC run are eluted

in less than three minutes (circled in green). Those in the standard HPLC

run are eluted much later (circled in red). These chromatograms represent

a calibration level of 500 pg/mL (ppt).

Figure 4: Calibration curve for the compound azoxystrobin. This calibration

curve covers the range from 0.5 to 1000 pg/mL (ppt)