Background Image
Table of Contents Table of Contents
Previous Page  32 / 302 Next Page
Information
Show Menu
Previous Page 32 / 302 Next Page
Page Background

Results and Discussion

The analyst can select in which area to begin working

(Figure 1). In this application note, the entire process will

be illustrated, from method development to reporting.

Method Development

The Method Development section of the software allows

the user to select the compounds that will be analyzed in

the method. In this experiment, the appropriate SRM

transitions for the triazine mixture were chosen from the

CDS and inserted into the method for detection (Figure 2).

No compound optimization is necessary for compounds

already in the data store.

Additionally, the calibration standards, QC levels, and

peak detection settings are defined in the Method

Development section. Results can be flagged based on

user-defined criteria. For example, the user can set a flag

for a compound whose calculated concentration is beyond

the upper limit of linearity, above a

defined reporting limit, or below a

limit of detection. This allows for

faster data review after collection,

and quick identification of positive

samples. Full support for qualifier

SRM ion ratios is also included but

was not used in this experiment.

Acquisition

The Acquisition section provides a

step-by-step process to acquire data.

The progress is followed in an

overview section on the left side of

the screen (Figure 3). A green

checkbox indicates that the step has

been completed and there are no

errors. The steps include template

selection (pre-defined sample lists,

which are helpful in routine analysis),

method selection, sample list

definition, report selection, and instrument status. Figure

3 shows calibrators, blanks, replicate “unknowns” of a

1 pg/mL sample, and drinking water samples for this

experiment.

A final status page summarizes the method and all of

the samples to be run and gives an overall summary of the

status of the instrument (Figure 4). Three color-coded dots

are shown: green indicates an ‘ok’ status; yellow indicates

the instrument module is in standby; and red indicates the

Figure 1. TraceFinder Welcome screen

Figure 2. Master Method View, showing the triazine compounds that will be monitored in this method.