Background Image
Table of Contents Table of Contents
Previous Page  34 / 302 Next Page
Information
Show Menu
Previous Page 34 / 302 Next Page
Page Background

instrument module is either off or disconnected. From the

final status page, the batch can be acquired or saved to be

run at a later date. A previously saved calibration curve

can be used, so that a calibration need not be run every

day. For example, the save function can be used to

prepare for future batches in advance of sample

preparation. When the samples are ready to be run, the

previously saved batch is loaded and acquisition is begun.

Data Review

The targeted analysis of triazine compounds in drinking

water samples was reviewed in the Data Review section of

TraceFinder. In this section, calibration lines, ion ratios,

peak integration, and mass spectra (if applicable) can be

monitored. In addition, the Data Review section can flag

samples that meet certain user-set criteria. For example, a

limit can be set on the R

2

value of a calibration line. A

green flag means that all user-set criteria have been met,

while a red flag indicates that the sample exceeds or fails

some user-set criteria and a yellow flag indicates that the

compound was not found in the sample. Flags can also be

used to highlight “positive” or “negative” hits in a

sample. Figure 5 illustrates the red flags indicating the

absence of peaks in blank samples for the compound

simazine at its lowest calibration level, 100 fg/mL. In

addition, flags can be set to alert for the presence of

carryover in blank samples. In this study, 20 mL injections

of the calibration standards, even at the highest level,

resulted in no detectable carryover.

The Data Review pane allows user adjustments, such

as peak reintegration. The effects of the changes on the

results are instantly updated in the results grid. Excellent

linearity was observed for all analytes, with R

2

values

ranging from 0.9921 for atrazine to 0.9995 for propazine

and terbuthlazine (co-eluting isomers, summed together

for this analysis).

As mentioned previously, no carryover was observed

in the blank samples, which illustrates the ability to use a

single loading column for multiple analyses of drinking

water samples. No triazines were detected in the soda

sample, but one of the commercial drinking water samples

tested positive for atrazine. The concentration of atrazine

in the sample was calculated to be 0.24 pg/mL, well below

the regulatory levels in the United States and Europe.

However, using standard injection techniques without

sample preconcentration, it is unlikely that this amount of

atrazine would be detected in a typical LC-MS/MS

analysis of triazines.

In addition to 20 mL injections, 1 mL and 5 mL

injections were analyzed in a separate experiment. The

%RSDs for replicate injections, without internal

standards, at 20 mL are shown with all of the compounds

in Table 1.

Figure 5. Data Review section. The red flags for blank samples indicate that peaks were not found in these samples.