Background Image
Table of Contents Table of Contents
Previous Page  261 / 302 Next Page
Information
Show Menu
Previous Page 261 / 302 Next Page
Page Background

Determination of 1,4-Dioxane in Drinking

Water by Gas Chromatography/Mass

Spectrometry (GC/MS) with Selected Ion

Monitoring (SIM)

Mark Belmont, David Steiniger, Eric Phillips, Sergio Guazzotti, Pat O’Brien, Alexander Semyonov

Thermo Fisher Scientific, Austin, TX

Application Note 52295

Key Words

ISQ Single Quadrupole GC-MS, TRACE GC Ultra, TriPlus RSH autosampler,

PTV inlet, Sequential SIM/Full Scan, EPA Method 522, Environmental

Introduction

1,4-Dioxane is used mainly as a stabilizer for

1,1,1-trichloroethane for transport in aluminum

containers. It is an irritant to eyes and respiratory system

and suspected of causing damage to nervous system, liver,

and kidneys.

1

In 2008, testing sponsored by the U.S.

Organic Consumers Association found dioxane in almost

half of tested organic personal-care products.

1

Of the total

1.163 million pounds of 1,4-dioxane released into the U.S.

environment in 1992, as reported to the Toxics Release

Inventory, 680 thousand pounds (58.5%) were released

into the atmosphere, 450 thousand pounds (38.7%) were

released into surface waters, and 33 hundred pounds

(2.8%) were released onto the land (TRI92 1994).

2

In

2005, the New Hampshire Department of Environmental

Services Waste Management Division started enforcement

of an Ambient Groundwater Quality Standard reporting

limit of 3 µg/L and trending towards a detection limit of

0.25 µg/L. 1,4-Dioxane has been detected in drinking

water in the U.S. at a concentration of 1 µg/L. This

application highlights the use of SIM/Full Scan to identify

unknowns with a NIST library, while producing accurate

results that meet EPA Method 522 requirements.

Experimental Conditions

Data was collected using a Thermo Scientific ISQ single

quadrupole mass spectrometer utilizing the Thermo

Scientific TriPlus RSH autosampler and a PTV inlet

(CT-Splitless mode) on a Thermo Scientific TRACE GC

Ultra gas chromatograph. The mass spectrometry data

was collected in Full Scan (FS), selected ion monitoring

(SIM), and SIM/Scan modes. A Thermo Scientific

TraceGOLD TG-624 column (30 m × 0.25 mm ID, 1.4 µm

film thickness; p/n 26085-3320) was used with a Siltek

®

deactivated baffle liner (p/n 453T2120). Table 1 lists the

GC parameters. The ion source temperature of the mass

spectrometer was set to 230 °C. The instrument was tuned

to meet the bromofluorobenzene (BFB) criteria for this

method. See Figure 1.

1,4-Dioxane calibration standards were prepared in

dichloromethane as per the method to provide a range

from 0.05 ppb to 40 ppb of dioxane.

Table 1. GC parameters

GC Oven Ramp

Ramp

Temp

Hold

30 ˚C

1 min

7 ˚C/min

90 ˚C

0 min

20 ˚C/min

200 ˚C

3 min

PTV Inlet

Temperature

200 ˚C

Split Flow

30 mL/min

Splitless Time

0.50 min

Solvent Valve Temp

100 ˚C