Application Note 618

AN63992-EN 1114S

In some cases, real samples may present concentrations up

to 30 mg/L. For this reason, it was decided to evaluate the

upper limit of quantitation. A 50 mg/L solution of MPA

was prepared in bovine serum albumin (Sigma-Aldrich).

This concentration in the linear range gave accuracy

within 100 ± 15% and a CV less than 15%.

To determine the best weighting factor, concentrations

were back-calculated and the model with the lowest total

bias across the concentration range was considered the

best suited. Four-point calibration standard curves were

calculated and fitted by linear models. To determine

LLOQ, a ten-fold dilution with bovine serum albumin

was made from Calibrator 1 to get a concentration of

0.097 mg/L. At this LLOQ, the accuracy and precision

values were, respectively, 108% and 4.6% for 10

replicates.

Figure 2 shows a representative calibration curve of MPA.

Figure 2. Calibration curve of MPA.

MPA

Y = 0.664727*X R^2 = 0.9986 W: Equal

0 1 2 3 4 5 6 7 8 9 10

0.0

0.5

1.0

1.5

2.0

2.5

3.0

3.5

4.0

4.5

5.0

5.5

6.0

6.5

Area Ratio

Carryover

Carryover was evaluated by injecting two levels of plasma

in triplicate: a blank plasma (B) and a high-level plasma

spiked at 20 mg/L (H). The injection sequence was three

injections of the high level (H

1

, H

2

, H

3

) followed by three

injections of the low level (B

1

, B

2

, B

3

). The sequence was

repeated five times and the carryover was calculated using

the following formula expressed as a percentage:

Carryover = (b1-b3) / (h-b3) x 100

b1: average concentration obtained for all B1

injections

b3: average concentration obtained for all B3

injections

h: average concentration obtained for all H

injections (from H1 to H3)

Carryover was evaluated to be less than 1.6%.

Conclusion

A simple and fast LC-MS/MS method was analytically

validated for the analysis of mycophenolic acid for

research purposes. Intraday and interday accuracy and

precision were successfully assessed in plasma-based

samples.

For Research Use Only. Not for use in diagnostic procedures.

Africa

+43 1 333 50 34 0

Australia

+61 3 9757 4300

Austria

+43 810 282 206

Belgium

+32 53 73 42 41

Canada

+1 800 530 8447

China

800 810 5118

(free call domestic)

400 650 5118

Denmark

+45 70 23 62 60

Europe-Other

+43 1 333 50 34 0

Finland

+358 9 3291 0200

France

+33 1 60 92 48 00

Germany

+49 6103 408 1014

India

+91 22 6742 9494

Italy

+39 02 950 591

Japan

+81 45 453 9100

Korea

+82 2 3420 8600

Latin America

+1 561 688 8700

Middle East

+43 1 333 50 34 0

Netherlands

+31 76 579 55 55

New Zealand

+64 9 980 6700

Norway

+46 8 556 468 00

Russia/CIS

+43 1 333 50 34 0

Singapore

+65 6289 1190

Spain

+34 914 845 965

Sweden

+46 8 556 468 00

Switzerland

+41 61 716 77 00

UK

+44 1442 233555

USA

+1 800 532 4752

Thermo Fisher Scientific,

San Jose, CA USA is

ISO 13485 Certified.

ISO13485

www.thermoscientific.com

©2014 Thermo Fisher Scientific Inc. All rights reserved. ISO is a trademark of the International Standards Organization.

MassTox is a registered trademark of ChromSystems Instruments & Chemicals GmbH. Sigma-Aldrich is a registered trademark

of Sigma-Aldrich Co. All other trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is

presented as an example of the capabilities of Thermo Fisher Scientific products. It is not intended to encourage use of these

products in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are

subject to change. Not all products are available in all countries. Please consult your local sales representative for details.