Background Image
Table of Contents Table of Contents
Previous Page  448 / 658 Next Page
Information
Show Menu
Previous Page 448 / 658 Next Page
Page Background

2

Veri cation of the Simultaneous Analysis of Heroin Addiction Treatment Compounds Using LC/MS/MS with a New Prelude SPLC

System

Overview

Purpose:

There are several compounds used for the treatment of heroin addiction.

These compounds include methadone, buprenorphine, norbuprenorphine, naloxone,

naltrexone , and their metabolites. The metabolites of interest are 2-ethylidine-1,5

dimethyl-3,3-diphenylpyrrolidine (methadone metabolite, aka EDDP), buprenorphine

glucuronide (buprenorphine metabolite), and norbuprenorphine glucuronide

(norbuprenorphine metabolite). All total, the analysis of these compounds for research

includes 8 analytes with 4 internal standards, that are commonly used in the treatment

of heroin addiction.

Methods:

Samples for this analysis were prepared in human urine. After the addition of

internal standard, they were injected for analysis using the

Thermo Scientific™

Prelude™ SPLC sample preparation

-liquid chromatography system. This system was

fitted with a Thermo Scientific™

Accucore

™ 100x3.0, 2.6 µm particle size column for

separation. Additionally, a Thermo Scientific™ TSQ Vantage™ mass spectrometer in

positive ion mode was used for analyte detection.

Results:

All 8 compounds were simultaneously verified using the Prelude SPLC

system and the TSQ Vantage MS. The resulting chromatography, correlation

coefficients, standard curve linearity, quality control data, and analyte transitions are

explained in the following sections to illustrate the success of this analysis.

Introduction

Several different compounds are currently used in the treatment of heroin addiction.

These compounds and their metabolites were analyzed for research using the new

Prelude SPLC system and a TSQ Vantage MS. This workflow takes advantage of a low

system volume to decrease solvent consumption and successfully quantify methadone,

2-ethylidine-1,5 dimethyl-3,3-diphenylpyrrolidine (EDDP), buprenorphine,

buprenorphine glucuronide, norbuprenorphine, norbuprenorphine glucuronide,

naloxone, and naltrexone. This work verifies a heroin treatment panel method

performed on the Prelude SPLC system. In order for this system to be evaluated, it

must fall within certain acceptance criteria. These set parameters are designed to

determine the success or failure of a particular LC-MS/MS workflow. These parameters

include, but are not limited to:

1) The lower limit of quantitation (LLOQ) and low quality control need to be

±

20% of

the expected concentration.

2) All of the remaining calibrators and controls need to be

±

15% in order for the

instrument to be successfully validated.

3) All of these requirements must be met for three consecutive days so that interday

and intraday accuracy and precision can be determined.

4) The signal in the blank following the highest standard may not exceed 20% of the

LLOQ signal. This factor is often called carryover.

Methods

Sample Preparation

Human urine was spiked with all 8 analytes and then serial diluted into a calibration

curve. Buprenorphine, norbuprenorphine, buprenorphine glucuronide, and

norbuprenorphine glucuronide had an analytical measurement range of 1.0 ng/mL to

100 ng/mL. Methadone, EDDP, naloxone, and naltrexone had an analytical

measurement range of 5.0 ng/mL to 500 ng/mL. Quality controls were also prepared in

human urine at three different levels. The urine aliquots were diluted with a combination

of water and methanol that contained internal standards. These samples are then

injected onto the system for analysis.

Liquid Chromatography

Chromatographic separations of all compounds were performed using Prelude SPLC

system, seen in Figure 1, equipped with an Accucore 100x3.0mm C18 analytical

column with 2.6 µm particle size. The system mobile phases consisted of 10mM

ammonium formate, 0.05% formic acid in water and 10mM ammonium formate, 0.05%

formic acid in methanol. The system needle washes were 60% water, 40% methanol,

and 0.5% formic (aqueous) and 45% isopropanol, 45% acetonitrile, and 10% acetone

(organic).

Mass Spectrometry and Data Analysis

The detector was a TSQ Vantage triple-stage quadrupole mass spectrometer with

HESI-II ionization probe in positive ion mode. Quantitation of results was performed

using Thermo Scientific™ LC

QUAN

™ software.

Results

Analyte result summa

Buprenorphine, norbupr

glucuronide were all pre

concentrations at 3.0, 4

naltrexone were prepar

concentrations of 15.0,

for each analyte. Metha

EDDP. Naloxone-d5 wa

Buprenorphine-d4 was

norbuprenorphine-d3 fo

transitions used for the

Compound

methadone

EDDP

naloxone

naltrexone

buprenorphine

norbuprenorphine

buprenorphine

glucuronide

norbuprenorphine

glucuronide

methadone-d9

naloxone-d5

buprenorphine-d4

norbuprenorphine-d3

FIGURE 1. Prelude SP

Table 1. Analyte Transi