Background Image
Table of Contents Table of Contents
Previous Page  232 / 302 Next Page
Information
Show Menu
Previous Page 232 / 302 Next Page
Page Background

2

Experimental

Sample Preparation

Preparation of standards

A methyl tertiary-butyl ether (MTBE) solution combining

four types of haloacetic acids [monochloroacetic acid

(MCAA), dichloroacetic acid (DCAA), trichloroacetic acid

(TCAA), and monobromoacetic acid (MBAA), 100 mg/L

of each, Kanto Kagaku] was used for the haloacetic acid

standard solution. The solution was diluted in ultrapure

water and used to prepare the calibration curve.

Preparation of laboratory fortified matrix

The following anions were added to a final concentration

as shown: Cl

-

, 35 mg/L; SO

4

2

-

, 35 mg/L; NO

3

-

, 50 mg/L.

Ascorbic acid was added at 10 mg/L level.

Preparation of the sample

After sampling the tap water, ascorbic acid was added at

the level of 10 mg/L for residual chlorine removal.

Liquid Chromatography

Equipment:

Thermo Scientific

Dionex

UltiMate

3000 RSLC system, which included

the LPG-3400RS Quaternary Rapid

Separation Pump, WPS-3000TRS Rapid

Separation Thermostatted Wellplate Sampler,

and TCC-3000RS Rapid Separation

Thermostatted Column Compartment

Column:

Acclaim HAA column

(2.1 x 50 mm, 3 µm), P/N SP6917

Mobile phase A:

Water (LC/MS grade)

Mobile phase B:

200 mM aqueous ammonium sulfate solution

Mobile phase C:

Acetonitrile

Gradient:

Refer to Figure 1

Flow rate:

0.3 mL/min

Operating temperature: 25.0 °C

Injection volume:

50 µL

Figure 1. LC gradient

To extend the life of the columns, they should be stored

in a 100 mM acetic ammonium (pH 5.0)/acetonitrile

(1:4 v/v) solution.

Mass Spectrometry

Equipment:

Thermo Scientific

TSQ Quantum Ultra

triple-stage quadrupole MS

Ionization method:

Negative ESI

Spray voltage:

500 V

Sheath gas:

60 arbitrary units

Aux gas:

10 arbitrary units

Capillary temperature:

250 °C

Vaporizer temperature: 400 °C

Skimmer offset:

10 V

Collision gas pressure: Ar, 0.8 mTorr

Cycle time:

2 ms

Mass resolution:

Q1: 1.5 Da (SRM mode)

SRM transitions:

Refer to Table 1

Results and Discussion

Separation of Matrix Ions

When optimizing separation conditions, it is important to

adequately separate matrix ions and haloacetic acids.

However, further care is required to separate chloride ions

and MCAA. Detection close to the MCAA retention time

was confirmed using accurate mass MS.

1

Cl

-

(

m/z

35) is

detected as fragment ions from NaCl

2

-

(

m/z

93) using

CID. It is therefore detected in the same transition as

MCAA. If the retention mechanism is unclear, or if

separation of MCAA and chloride cannot be confirmed,

false quantification could result, depending on the

behavior of the chloride ions. This is why processing to

remove chloride ions is recommended in analysis systems

using ODS columns.

In this investigation for the Acclaim HAA column, the

resolution of HAAs from interfering anions in a synthetic

sample matrix spiked with HAAs is demonstrated

(Figure 2).

Compound

Name

Precursor

(

m/z

)

Product

(

m/z

)

CE

(eV)

MCAA quantitation ion

93

35

10

MCAA qualifying ion

95

37

10

DCAA quantitation ion

127

83

10

DCAA qualifying ion

129

85

10

TCAA quantitation ion

161

117

10

TCAA qualifying ion

163

119

10

Table 1. SRM transitions