AI10382-GC-MS-Food Safety-Analysis - page 172

AN20734_E 05/13S
Application Note 20734
thermoscientific.com/columnsforgc
© 2013 Thermo Fisher Scientific Inc. All rights reserved. Vespel is a registered trademark of E. I. du Pont de Nemours and Company. All other trademarks are
the property of Thermo Fisher Scientific Inc. and its subsidiaries. This information is presented as an example of the capabilities of Thermo Fisher Scientific Inc.
products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Specifications, terms
and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details.
USA and Canada
+1 800 332 3331
France
+33 (0)1 60 92 48 34
Germany
+49 (0) 2423 9431 20 or 21
United Kingdom
+44 (0)1928 534110
Japan
+81 3 5826 1615
China
+86 21 68654588 +86 10 84193588
+86 20 83145199 800 810 5118
India
+91 22 6742 9494 +91 27 1766 2352
Australia
1 300 735 292 (free call domestic)
New Zealand
0800 933 966 (free call domestic)
All Other Enquiries
+44 (0) 1928 534 050
Technical Support
North America
+1 800 332 3331
Outside North America
+44 (0) 1928 534 440
Conclusion
HyperSep Hypercarb SPE cartridges offer high levels of reproducibility as well as cleaner extracts,
which yields very good results. TraceGOLD TG-WaxMS GC columns are suitable for the GC-MS
analysis of acrylamide because of the low bleed stationary phase and better retention of polar
analytes compared with lower polarity stationary phases.
Reference
[1] Acrylamide in Drinking-water - Background document for development of WHO Guidelines
for Drinking-water, 2011,
/
acrylamide.pdf
Specified
Concentration
(µg/mL)
Calculated Concentration % Difference
0.25
0.225
-9.83
0.50
0.469
-6.22
1.00
0.983
-1.70
2.50
2.520
0.79
5.00
4.979
-0.41
10.0
10.037
0.37
Table 1: Accuracy data for the standard addition calibration curve for spiked
acrylamide in potato chips
Figure 1: TIC of chromatogram of 1 µg/mL spiked acrylamide (
m/z
71) and acrylamide-d
3
(
m/z
74) extracted from
potato chips
Time (min)
9.0
10
0
20
30
40
50
60
70
80
90
9.5
10.0
10.5
11.0
100
Relative Abundance
10
0
20
30
40
50
60
70
80
90
100
Acrylamide
Acrylamide-d
10.05 min
Impurity
Impurity
3
1...,162,163,164,165,166,167,168,169,170,171 173,174,175,176,177,178,179,180,181,182,...225