          2
        
        
          Experimental Details
        
        
          Consumables
        
        
          Part Number
        
        
          Column:
        
        
          TRACE TR-FAME GC column,
        
        
          260M238P
        
        
          100 m × 0.25 mm × 0.20 μm
        
        
          Septum:
        
        
          Thermo Scientific BTO, 17 mm
        
        
          31303211
        
        
          Liner:
        
        
          Thermo Scientific™ Split FocusLiner™, 3 × 8 × 105 mm 45350031
        
        
          Column ferrules:
        
        
          100% graphite ferrules for Thermo Scientific™ 29053488
        
        
          TRACE™ injector 0.1–0.25 mm i.d.
        
        
          Column ferrules:
        
        
          Graphite/Vespel
        
        
          ®
        
        
          for transfer line 0.1–0.25 mm i.d.
        
        
          29033496
        
        
          Injection syringe:
        
        
          10 μL fixed needle syringe for Thermo Scientific™
        
        
          36500525
        
        
          TriPlus™ Autosampler
        
        
          Vials and closures:
        
        
          Thermo Scientific™ Chromacol™ 9 mm screw 0.3 mL 03-FISV (A)
        
        
          fixed insert amber Micro+ vials
        
        
          Chromacol 9 mm screw caps with Silicone/PTFE
        
        
          9-SC(B)-ST101
        
        
          septa
        
        
          Syringes:
        
        
          Thermo Scientific™ National™ 30 mm GMF Syringe
        
        
          F2500-20
        
        
          filter membrane, 3.1 µm pore size
        
        
          Thermo Scientific™ National™ Target™ 3 mL plastic
        
        
          S7510-3
        
        
          disposable syringes
        
        
          Sample Handling Equipment
        
        
          Part Number
        
        
          Thermo Scientific™ Reacti-Therm™ III Heating/Stirring Module
        
        
          TS-18823
        
        
          Thermo Scientific™ Reacti-Vap™ III Evaporator
        
        
          TS-18826
        
        
          Thermo Scientific Reacti-Vap Block
        
        
          TS-18814
        
        
          Thermo Scientific™ Reacti-Vial™ Reaction Vials 10 mL
        
        
          TS-13225
        
        
          Chemicals and Reagents
        
        
          Part Number
        
        
          Fisher Scientific™ HPLC grade hexane
        
        
          H/0403/15
        
        
          Fisher Scientific HPLC grade water
        
        
          W/0106/17
        
        
          Fisher Scientific HPLC grade methanol
        
        
          M/4056/17
        
        
          Fisher Scientific HPLC grade potassium hydroxide
        
        
          S/9220/PB08
        
        
          Thermo Scientific 14% BF
        
        
          3
        
        
          -methanol esterification reagent
        
        
          TS-49370
        
        
          Sample Preparation
        
        
          Butter, margarine, and palm oil were treated by two derivatization methods. In the first method, potassium hydroxide /
        
        
          methanol was used to esterify the fat samples [1]. This was then compared to an acidic derivatization method involving
        
        
          BF
        
        
          3
        
        
          -methanol [2].
        
        
          Base esterification:
        
        
          A 50 mg liquid fat sample was weighed into Reacti-Vial containing
        
        
          a magnetic stirrer and 1 mL of hexane and 2 mL of 4 mol/L potassium
        
        
          hydroxide / methanol were added. The Reacti-Vial was capped and
        
        
          placed in the Reacti-Therm module for 30 minutes at 50 °C. The
        
        
          mixture was cooled to room temperature and 1 mL of water was then
        
        
          added. After phase separation, an aliquot of the organic layer was
        
        
          transferred to a fixed insert GC vial.
        
        
          Acid esterification:
        
        
          A 50 mg liquid fat sample was weighed into a Reacti-Vial containing
        
        
          a magnetic stirrer and 1 mL of hexane and 0.5 mL of Thermo Scientific
        
        
          14% BF
        
        
          3
        
        
          -methanol was added. The Reacti-Vial was capped and
        
        
          placed in the Reacti-Therm module for 30 minutes at 50 °C. The
        
        
          mixture was cooled to room temperature and 1 mL of water was then
        
        
          added. After phase separation, an aliquot of the organic layer was
        
        
          transferred to a fixed insert GC vial.