Background Image
Table of Contents Table of Contents
Previous Page  74 / 302 Next Page
Information
Show Menu
Previous Page 74 / 302 Next Page
Page Background

The high-purity IC eluent is automatically produced in

situ (Figure 1). The pump delivers water to an eluent

generator cartridge (EGC) that converts the water into the

selected concentration of potassium hydroxide eluent

using electrolysis. After separation on the column, the

eluent enters the anion self-regenerating suppressor

(ASRS) that produces hydronium ions to exchange with

potassium in the eluent. This makes the mobile phase

compatible with the mass spectrometer liquid inlet system.

Mass Spectrometry

MS analysis was carried out on a Thermo Scientific TSQ

Quantum Access triple stage quadrupole mass

spectrometer with an electrospray ionization (ESI) source.

The MS conditions were as follows:

Ion source polarity:

Negative ion mode

Spray voltage:

3000 V

Sheath gas pressure:

40 arbitrary units

Ion sweep gas pressure: 1 arbitrary unit

Auxiliary gas pressure: 2 arbitrary units

Capillary temperature: 400 °C

Collision gas pressure: 1.5 bar

Scan conditions:

Table 1

Table 1. MS Scan Conditions

Scan

Scan

Collision

Tube

Compound

Mass

Width Time (s)

Energy

Lens

AMPA

110.17 / 63.3

0.01

0.5

19

60

AMPA

110.17 / 79.2

0.01

0.5

35

60

Glyphosate

168.09 / 150.1

0.01

0.5

13

51

Glyphosate

168.09 / 79.4

0.01

0.5

40

51

Results and Discussion

The major matrix peaks of chloride, nitrate, carbonate,

and sulfate were well-resolved. The separation of all

compounds occurred in both dimensions in 30 minutes.

The calibration curves showed excellent linearity using

only external quantitative measurements without internal

standards.

For quantitation, samples were run in the MS/MS

selective reaction monitoring (SRM) mode on the TSQ

Quantum Access

triple stage quadrupole instrument. The

calibration range was 0.1–50 ppb for AMPA and

0.05–50 ppb for glyphosate. The correlation coefficients

(R

2

) of the 110

63 and 110

79 SRM transitions of

AMPA were both 0.9997 (Figures 2 and 3). The 168

150 transition of glyphosate had an R

2

of 0.9997 and the

168

79 transition yielded an R

2

of 0.996 (Figures 4 and

5, respectively).

The minimum detection limit (MDL) in matrix was

calculated by seven replicate injections of 5 ppb in a

simulated matrix with high concentrations of chloride,

carbonate, nitrate, and sulfate (250 ppm chloride and

sulfate, 150 ppm sodium bicarbonate, 20 ppm nitrate).

The MDLs for AMPA and glyphosate were calculated by

using the equation MDL=

t

99%

× S

(n-1)

, where

t

equals the

Student’s

t

test at 99% confidence intervals (

t

99%, (6)

=

3.143) and S is the standard deviation. The calculated

MDL for AMPA was 0.313 ppb and 0.252 ppb for

glyphosate. This is well below the current MDLs of 6 ppb

for glyphosate in reagent water and 8.99 ppb in ground

water specified by the EPA guidelines found in Method

547.

Using ion chromatography to quantitate AMPA and

glyphosate accurately at this level without sample

pretreatment requires the use of a mass spectrometer.

However, the source of the instrument can be subject to

fouling from routine analysis of samples of high-ionic

strength. The use of multi-dimensional chromatography

significantly reduces the introduction of matrix ions to the

mass spectrometer, increasing the method robustness in

challenging sample matrices. The recoveries for AMPA

and glyphosate were 97.2% and 82.1%, respectively, for

5 ppb spiked into high-ionic strength samples. The relative

standard deviations were less than 5% for both

compounds, even without an internal standard (Figure 6).

Figure 1. The flow schematic for a two-dimensional IC-MS/MS application. The first dimension separates the analytes of interest from a majority of the

matrix ions. The second dimension improves peak shape and keeps the source of the MS clean.