phase (5% diphenyl/95% dimethylpolysiloxane, 30

m x 0.25 mm ID, 0.25 µm film thickness, e.g. Thermo

Scientific™ TraceGOLD™ TG-5MS p/n 26098-1420).

The TG-624 column type is a mid-polarity column

ideally suited for the analysis of volatile analytes, whereas

3

Injector, temperature prog.

PTV-LVI

40 °C, 0.1 min (injection phase, @ 100 kPa)

10 °C/min to 80 °C, 0.3 min (@ 200 kPa)

10 °C/min to 110 °C (transfer phase)

14.5 °C/min to 290 °C (cleaning phase)

Split flow

20 mL/min

Solvent vent

open until 0.17 min

closed until 4.17 min

open until end of run

Injection mode, volume

split, 4 µL

Carrier gas, flow

Helium, constant flow 1 mL/min

Oven program

40 °C, 5 min

40 °C/min to 200 °C

200 °C, 5 min

Transfer line temperature

205 °C

Table 1

.

GC Conditions

Ionization

EI, 70 eV

Scan mode, range

SIM,

m/z

76, 78

Acquisition rate

2 scans/s

Ion source temperature

200 °C (optimized for CS

2

S/N ratio)

Table 2. MS Conditions

Figure 3. CS

2

chromatogram, 5 ppb matrix spike calibration.

the TG-5MS column is more commonly used especially

for pesticide analysis and is commonly available in all

laboratories. Both columns were thus tested for the

applicability of the method. Either column can be used

for the DTC analysis.